Operating Manual for Euroklav
®
23V-S
Page 37
Warning message
Cause / Remedy
The pressure for the ventilation drying lies outside the
permitted range. The report comes at the end of the program,
and as the last line of the print-out:
The sterile filter may be clogged or torn. Exchange the
sterile filter (MELAG Art. No.: 53390).
Communication with the printer via the serial interface has
been interrupted. This message appears when a report cannot
be printed out. It is displayed for 20 seconds. If the printer
becomes operational during this period the cycle record prints
out:
•
The autoclave may be operated without a printer. Check
under the "Data transfer" menu that the option "No printer"
has been selected. (see Section 6.3.1.3)
•
Check the cable connection between the printer and the
autoclave.
•
Check the power supply to the printer. In the MELA
®
42
the red light should indicate 'power on'
•
The printer may be "Offline". Select "online"
(MELA
®
42, press "SEL" button, green LED "SEL"
should shine)
The internal printer memory is full (40 cycles recorded), an
external printer is registered, and in the "Print" menu the
option "Immed. print-out? No" is selected. The message is
displayed when a program is started. Pressing the "Start /
Stop" button again deleted the message and the program
starts:
•
You can continue operations simply by pressing the "Start /
Stop" button twice when you start a program.
•
Select "Immed. print-out? Yes" (see Section 6.3.1.3)
•
Delete stored records (See Section 6.3.6), if necessary
print-out all stored cycle records first (see Section 6.3.4)
In the Data transfer menu, select the "No printer" option (see
Section 6.3.1.3)
The service message is activated after a certain number of
loads or a set operating period, when a service is due. The
message appears before the start of every program. If you
press the "Start / Stop" button again the message is deleted
and the program starts.
•
You can continue operations, by simply pressing the
"Start / Stop" button twice when you start a program.
•
Have a service carried out as recommended by an
authorised MELAG servicing company or your specialist
dealer.
The cycle counter for servicing should be reset after the
service.
The leak rate during the vacuum test exceeds the limit value:
Check and if necessary clean the door seal and the rim of
the chamber
Repeat the vacuum test with a completely cold autoclave
If no other error reports occur during operation, you can
continue to use the autoclave until the regular service, when
the cause of the leak will be identified.
A c k n o w l e d g e
w i t h b u t t o n " - "
S t e r i l e f i l t e r
R e p l a c e !
W A R N I N G !
P r i n t e r m e m o r y
f u l l
P r i n t e r i s
N o t r e a d y
P l e a s e s e r v i c e
T e s t u n s u c c e s s f u l
L e a k r a t e : 3 . 8