Operating Manual for Euroklav
®
23V-S
Page 19
4.12.2 Terminate
drying _________________________________________________________________
A program can also be terminated during the drying phase, Since in this case the sterilization has been
completed, the items in this case can be treated as sterilized.
However, depending on the stage at which the drying program is interrupted, the load may not have dried
sufficiently, and wrapped items in particular may not be dry enough for sterile storage. We therefore
recommend that you do not interrupt the drying process for wrapped items in the "Universal Program" or
"Gentle Program".
With the "Quick Program" it may be desirable to interrupt the drying program so that items can be used
again. The unwrapped items will dry as they are cooling down.
Warning!
If the drying process is interrupted than steam may be released when the door of the autoclave is
opened.
Operation
Display message
1. The autoclave is in the drying phase. The
display shows the drying time
alternately with
the option to terminate the drying phase
2. Press the "Start-Stop" button
To confirm, press the "Start-Stop" once
again within 5 secs.
If no confirmation is given then the program
resumes normally.
3. If the "Start-Stop" button has been pressed
again to confirm then the program
terminates.
4. After the ventilation of the chamber the
display shows that the program has been
successfully completed.:
alternately with:
If a printer is connected and an immediate
report has been selected, this is printed-out,
together with confirmation that the drying
process has been terminated.
Program Start-Stop
I m m e d i a t e r e m o v a l
P r e s s ' S T O P '
Program Start-Stop
V a c u u m - d r y i n g
s i n . 3 ' – 0 . 9 b a r 6 8 ° C
Program Start-Stop
S t o p p r o g r a m ?
P r e s s ' S T O P '
Program Start-Stop
D r y i n g
s t o p p e d
Program Start-Stop
O p e n d o o r
p l e a s e
Program Start-Stop
Q u i c k p r o g r a m
r u n s u c c e s f u l l y
Program Start-Stop
D r y i n g
s t o p p e d