MELAG Euroklav 23V S Скачать руководство пользователя страница 12

 

 

Page 12 

4.3 

Loading the autoclave

_______________________________________________

 

It is of crucial importance for effective sterilization and good drying that the autoclave is loaded properly:  
When loading the autoclave, take account of the following points:  

Tray rack  

For the Euroklav

®

23V-S there are 2 types of tray rack:  

Type "B" (MELAG-Art.-No.: 40224) for up to 4 trays or 4 standard tray-cassettes 
Type "C" (MELAG-Art.-No.: 40242) for up to 6 trays or 3 standard tray-cassettes 
Both  types of tray rack are also suitable for the MELAG-sterilization containers Type 15K,M,G; Type 
17K,M,G,R; Type 23R,M,G, Type 28M,G).  
Normally, the autoclave should be used in conjunction with a tray rack, since this ensures that steam 
penetration and drying are as good as possible. In exceptional situations (e.g. when using sterilization 
containers from other manufacturers), and after consultation with your specialist dealer or with MELAG, the 
tray rack can be removed and the container can be placed directly in the autoclave chamber.  
For the sterilization of instruments sealed in  transparent sterilization wrapping, it is recommended that you 
use the foil stand MELAG-Art.-No.: 22420. This contributes considerably to the drying process for such 
wrapped instruments. 

Trays 

Trays for objects which are to be sterilized must be perforated, in order to allow condensation to run away. 
MELAG-trays are recommended. If you use dishes or trays without perforations, then the objects being 
sterilized will not dry properly. 

Enclosed sterilization containers 

Enclosed sterilization containers must be perforated on at least one side (preferably underneath) or must 
have valves, in order to ensure that steam can penetrate and condensate can run out. All MELAG-
sterilization containers meet these requirements with perforations on two sides and filter-cloth- inlays. 
Sterilization containers which only have perforations on the top only allow limited drying.  
If sterilization containers are stacked in the autoclave, it is important to ensure that the perforations are not 
blocked. 

Transparent sterilization packaging 

If you use transparent sterilization packaging, such as MELA

fol

® 

, then the items should if possible be stood 

vertically on the tray, or sterilized in foil holders (MELAG-Art.-No.: 22420). They should never be laid flat one 
on top of the other.  
If seals split open during sterilization it may be necessary to increase the length of the impulse on the sealing 
device or to use a double-seal.  
Standard tray- cassettes sealed in MELA

fol

®

 (250 mm wide) must be taped and clasped additionally to 

ensure that the side-seals do not split open. 

Multiple wrapping 

The pulsed vacuum method means it is possible to use multiple wrapping.  

Maximum loads  

Loads should not exceed 4 kg of instruments or 1 kg of textiles. 

Mixed loads 

If mixed loads of textiles and instruments are to be sterilized, then as far as possible the textiles should be 
above the instruments and direct contact with the instruments should be avoided.  
Inclusion of textiles and instruments in the same sterilization container is not desirable. 
Textiles should never come into direct contact with the walls of the chamber. 
If different types of packaging are included in a load, then: 

  Instruments and sterilization containers should be at the bottom  

  Transparent and paper sterilization packaging should be at the top (but lower than textiles)  

Liquids  

The appliance is not suitable for the sterilization of liquids! 

Suitability for sterilization  

Relevant information provided by manufacturers of instruments and textiles about sterilization should be 
strictly observed. 

Содержание Euroklav 23V S

Страница 1: ...ization equipment More than 335 000 MELAG units sold throughout the world testify to the exceptional quality of our sterilizers which are manufactured exclusively in Germany As all other MELAG product...

Страница 2: ...for one way water 8 2 6 Internal water supply with demineralized distilled water 8 2 7 Connection of a water purification system 8 3 INITIAL START UP 8 3 1 Printer connection Initialization optional...

Страница 3: ...clave malfunctions 34 7 2 Malfunctions without display messages 34 7 2 1 No display 34 7 2 2 Excessive water consumption 34 7 2 3 Poor drying 34 7 3 Warning messages 35 7 4 Error reports 38 8 TAKING C...

Страница 4: ...filter 10 Power switch 11 Overheating trip switch steam generator 3 Connection for internal distilled demineralized water supply 12 Sliding door lock 4 One way water outlet 3 4 13 Door left side hing...

Страница 5: ...3 Integrated steam generation ________________________________________________________ The powerful steam generation in the sterilization chamber makes it possible to sterilize large loads of instrum...

Страница 6: ...ecords from the memory 2 Installation When setting up and installing the autoclave please consult the separate instruction leaflet Installing the Euroklav 23V S 2 1 Setting up the autoclave __________...

Страница 7: ...checked with a spirit level at the chamber flange and then the front feet should be extended by giving them five 5 turns 5 3 2 9 7 8 1 6 4 Rear view Fig 3 Installation of the Euroklav 23V S with wall...

Страница 8: ...mark 2 7 Connection of a water purification system _______________________________ The autoclave can be connected directly to a water purification unit which provides the demineralized or distilled wa...

Страница 9: ...vacuation pressure has been reached The equalisation period begins 6 After waiting for the equalisation 5 min the measuring period starts running here e g for 7 min 52s 7 After the measuring period 10...

Страница 10: ...y instructions __________________________________________________ When opening the door particularly after interrupting the drying process residual steam can escape from the autoclave chamber After op...

Страница 11: ...to the autoclave on instruments film rust Even top quality stainless steel instruments can form rust if they are not handled properly e g if they are treated with the wrong chemical cleaning or disinf...

Страница 12: ...ate can run out All MELAG sterilization containers meet these requirements with perforations on two sides and filter cloth inlays Sterilization containers which only have perforations on the top only...

Страница 13: ...ation time 15 min Sterilization of all types of wrapped items except Hollow A in particular large amounts of textiles or thermolabile materials plastic rubber or mixed loads wrapped unwrapped Prion Pr...

Страница 14: ...in chamber and steam temperature 2 Phase 1 and following Depending on the program chosen and the temperature of the chamber at the start of the program a number of cycles of steam inflow and outflow f...

Страница 15: ...report since the sterilization itself has been completed However with the exception of the Quick Program you should normally wait until the drying is completed 8 Program end After ventilation of the c...

Страница 16: ...00 Pre vacuum Evacuation 0 77 65 2 02 19 Steam entry 0 01 99 0 03 16 1 Fractionation Steam entry 1 01 119 9 04 18 Press release 0 20 104 8 04 49 2 Fractionation Steam entry 1 01 120 0 05 52 Press rel...

Страница 17: ...tandard DIN 58953 Part 7 Section 7 contains the following comment about residual moisture on paper wrapping or transparent sterilization paper after sterilization small amounts of water on the wrappin...

Страница 18: ...s the Start Stop once again within 5 seconds If no confirmation is given then the program resumes normally 2 If confirmation is given then the program stops The pressure inside the autoclave will then...

Страница 19: ...may be released when the door of the autoclave is opened Operation Display message 1 The autoclave is in the drying phase The display shows the drying time alternately with the option to terminate th...

Страница 20: ...ch warning message occurs then you should consult Section 7 which provides detailed advice and possible operational errors 4 14 Operational pauses __________________________________________________ In...

Страница 21: ...s can be downloaded at a later stage via the serial interface RS232 When the memory is full 40 program runs then before the start of the next run the oldest record will automatically be overwritten If...

Страница 22: ...____________________________________________________ 6 3 1 2 1 Installation_____________________________________________________________________ Records and archives can also be kept by using an exter...

Страница 23: ...Hold down button and also press button Select Function menu sub menu Print 2 Press Program button Select Print menu sub menu File transmission 3 Press button select sub menu Immed print out Display s...

Страница 24: ...ress the or button until the right number is reached e g here No 25 6 Press Program button to start the print out of the selected record or to terminate press Start Stop and return to the Function men...

Страница 25: ...Press Program button to start the printout of all the stored records up to 40 or to terminate press Start Stop and return to the Function menu Once printing has started termination is only possible by...

Страница 26: ...ct Function menu submenu Print 2 Press Program button Select Print menu submenu File transmission 3 Press or button until the display shows the memory status e g 4 Press the Start Stop to return to th...

Страница 27: ...menu submenu Print 2 Press Program button Select Print menu submenu Data transfer 3 Press or until the display shows All cycles delete 4 Press the Program button to delete all records or press Start S...

Страница 28: ...tton Select Print menu submenu File transmission 3 Press or until the display shows Test printout 4 Then press the Program button for a test print out or press Start Stop to terminate 5 Then press Sta...

Страница 29: ...clock here for example 17 00 4 by pressing the or button the following options can be selected 5 After finding the required option e g Minute press the Program button and the current value flashes 6 P...

Страница 30: ...s switched on In the default setting on delivery the automatic preheating is on The current setting for the automatic preheating can be changed as follows Operation Display message 1 Hold down button...

Страница 31: ...lect Function menu submenu Print 2 Press or until the display shows the submenu Sum batch number 3 Press Program the display shows the current total load count e g 4 Close by pressing Start Stop and r...

Страница 32: ...s Program button the display will show the current option here intern 4 Pressing the Program button switches between the options internal and extern press again to return 5 When the correct option has...

Страница 33: ...d press the Start Stop button to return to the Function menu and 6 Press the Start Stop button again to return to the initial position 6 9 Program modifications _______________________________________...

Страница 34: ...nside of the door lining then replace the screw cap and reconnect the autoclave to the power supply If there is still no display when the autoclave is switched on or if the display blacks out repeated...

Страница 35: ...o external IN this case the water will be supplied from an external water purifier The internal flow monitor for the demineralized distilled water supply has not reacted when filling the steam generat...

Страница 36: ...he internal storage tank clean tank with distilled demineralized water and refill to max with purified water to specifications Water from the MELAdem 47 The demineralization cartridge in the ion excha...

Страница 37: ...on again deleted the message and the program starts You can continue operations simply by pressing the Start Stop button twice when you start a program Select Immed print out Yes see Section 6 3 1 3 D...

Страница 38: ...re release or for reaching the minimum drying pressure has been exceeded Check that the door seal and the lip of the opening to the pressure chamber are intact and clean Check that the autoclave is st...

Страница 39: ...34 The internal testing of the sensors for temperature pressure or conductivity showed an excessive deviation the error can be reported on switching on the appliance or in the course of a program If...

Страница 40: ...exceeded If this occurs repeatedly inform your specialist dealer The minimum sterilization temperature has not been reached Reduce the size of the load If this occurs repeatedly inform your specialist...

Страница 41: ...rder to avoid dirt and contamination being separated from the load in the autoclave and clogging filters valves and nozzles In particular locks joints and hinges must be cleaned thoroughly with a brus...

Страница 42: ...ily residues may have to be removed using a little washing up liquid and warm mains supply water with a suitable soft brush followed by swilling with distilled demineralized water Should the right tan...

Страница 43: ...out at the place of installation e g at 6 monthly intervals They shall demonstrate that sterilization is carried out satisfactorily when the operating instruction for the small sterilizer are followed...

Страница 44: ...le 1 Example 2 Euroklav 23V S rear view basic version Euroklav 23V S rear view with one way water Water supply be means of internal double chamber storage tank outflow installed Double chamber trap Ex...

Страница 45: ...2 3 1 Programs Operation times ______________________________________________________ Program Operation time without drying Drying time Warm start low load Warm max load Quick program unwrapped 134 C...

Страница 46: ...ds to steam condensing on the instruments and containers Formation of condensation on the sterilization container The steam also heats the objects contained in the sterilization containers Condensatio...

Страница 47: ...textiles on top of each other as this hinders the penetration of steam into the packages of textiles Loading textiles properly When loading sterilization containers with textile items care should be t...

Страница 48: ...as possible not be sterilized in the same load However where this is unavoidable for economic or other reasons the following rules should be observed Instruments and sterilization containers should b...

Страница 49: ...remain on the sterilized items However heat transfer from the sterilized objects can evaporate this after the sterilization process has been completed The German standard DIN 58953 Part 7 Section 7 co...

Отзывы: