Menu "Switch Unit
→
SCU Cntl"
OK
OK
SCU State
-> SCU Cntl <
SCU INFO
-> MANUAL: disabl <-
Selected Clk:2
SCU Outputs:
* enabled
* disabled
MANUAL:
Enabled | Disabled Switches between automatic and manual mode
Selected Clk:
1/2
Selects the master reference clock
3. Clock 1 / Clock 2 Switch (SDI 1 / SDI 2)
This switch is used to manually select the receiver or input reference in "Manual" mode. No automatic switchover
is performed in the event of the selected input reference losing its signal.
4. ACO Button (Access Control Override)
The button labeled ACO can be used if the access password has been forgotten. However, it will only work if
there is no connection with the "Meinberg Device Manager" software. If it is held down for a short time (for
around four seconds), the set password will be reset for 30 seconds to a blank field (i.e., you can simply press
RETURN when prompted for the password). After these 30 seconds, the system will then restore the password
to its previous value. This applies to connections made via both Telnet and mbgdevman. If a connection to
mbgdevman is established during the 30 seconds, the "empty" password will remain valid for the duration of this
connection (even if this connection is upheld for longer than 30 seconds). In this case, the old password will
only restored when the connection is terminated.
5. Network Interface (MDU only)
RSC switching modules used in an MDU system feature a network interface which can be used to configure the
system and components of the system; for this purpose you will require the "Meinberg Device Manager" software.
Download and documentation:
https://www.meinbergglobal.com/english/sw/mbg-devman.htm
8
Date: December 17, 2021
IMS-RSC180