www.megger.com
PAT400 Series
87
Appendix C - Specification
SPECIFICATIONS
PAT410
PAT420
PAT450
Electrical supply range
PAT-UK
230V ±10% ±1V @ 50Hz ±10% ±0.1Hz
Y
Y
Y
PAT-UK
110V ±10% ±1V @ 50Hz ±10% ±0.1Hz
Y
Y
PAT-DE
230V ±10% ±1V @ 50Hz ±10% ±0.1Hz
Y
Y
Y
PAT-CH
230V ±10% ±1V @ 50Hz ±10% ±0.1Hz
Y
Y
Y
PAT-EU
230V ±10% ±1V @ 50Hz ±10% ±0.1Hz
Y
Y
Y
PAT-NL/ES
230V ±10% ±1V @ 50Hz ±10% ±0.1Hz
Y
Y
Y
PAT-BE/FR
230V ±10% ±1V @ 50Hz ±10% ±0.1Hz
Y
Y
Y
PAT-PL
230V ±10% ±1V @ 50Hz ±10% ±0.1Hz
Y
Y
Y
PAT-IL
230V ±10% ±1V @ 50Hz ±10% ±0.1Hz
Y
Y
Y
PAT-AU
230V ±10% ±1V @ 50Hz ±10% ±0.1Hz
Y
Y
Y
PAT-US
120V ±10% ±1V @ 60Hz ±10% ±0.1Hz
Y
Instrument accuracy cannot be guaranteed when operated on supply frequencies other than those
specified above
Bond test (10A & 25A)
Y
Y
Open circuit voltage:
9 V a.c. ± 10% ± 0.1 V (supply 230V 50Hz)
Y
Y
Earth bond resistance accuracy:
±5% ±3 digits (0 to 0.49 Ω)
±5% ±5 digits (0.5 to 1.99 Ω)
Y
Y
Earth bond resistance resolution:
10 mΩ (0 to 1.99 Ω)
Y
Y
Display range:
0 to 1.99 Ω
Y
Y
Current Output
25A rms +4% -20% ±0.5A into 0.1Ω (230V 50Hz)
10A rms ±25% into 0.1Ω (230V 50Hz)
10A rms ±25% ±0.5A into 0.1Ω (120V 60Hz)
Y
Y
Y
Y
Y
Bond lead null range:
0.00 to 1.00Ω
Y
Y
Continuity test (200mA)
■
■
■
Continuity test voltage:
4.0 V d.c. -0% +10%
■
■
■
Continuity resistance accuracy:
±5% ±3 digits (0 to 0.49 Ω)
±5% ±5 digits (0.5 to 19.99 Ω)
■
■
■
■
■
■
Continuity resistance resolution:
10 mΩ (1 to 19.99 Ω)
■
■
■
Display range:
0.00 to 19.99 Ω
■
■
■
Current Output
200mA rms -0 +10% ±5mA into 2Ω
■
■
■
Continuity lead null range
0.00 to 9.99Ω
Y
Y
Y
Insulation test (250V & 500V)
■
■
■
Insulation resistance accuracy
(230V supply):
±2% ±5 digits (0 to 19.99 MΩ)
±5% ±10 digits (20 to 99.99 MΩ)
■
■
■
■
■
■
Insulation resistance resolution:
0.01 MΩ
■
■
■
Display range:
0.10 to 99.99 MΩ
■
■
■
Substitute leakage test
■
■
■
Test voltage and frequency
40 V a.c. ±10% @ nominal mains frequency
■
■
■
Leakage current accuracy:
±5% ±5 digits
■
■
■
Leakage current resolution:
0.01 mA
■
■
■
Display range:
0.00 to 19.99 mA
■
■
■
Reading corrected to nominal supply voltage
■
■
■