4 FUNCTIONS AND SET UP
4.7 Pass/Fail – P/F
In the P/F position you can enable and set the limit for
the Pass/Fail function. If the measured value exceeds
the set limit it will result in a notification on the dis-
play and by an audio signal
SET / P/F
Display abbreviations
SEL
Select
oFF
OFF
Set-up
1]
Select
SET / P/F.
Present setting is displayed.
2]
Press and hold OK key until character, "<" or
"oFF" start to blink.
3]
Select "oFF" or a value using the
▲ ▼
keys.
4]
Use the keys
▶
◀
to select the digit to
change, it will blink.
5]
Use the keys
▲ ▼
keys to set the desired
value.
6]
Press OK key.
To cancel press C key or turn the function
selector to another position.
4.8 PC communication – PC
COM
The PC COM position is used for all operations per-
formed from a PC using MOM2 Win. you can:
▪
Dump measurement data from the MOM2
▪
Delete all measurements in the data log
▪
Program "User" positions
▪
Set MOM2 clock
▪
Calibrate
PC COM
Display abbreviations
rEn
Remote
Set-up
The following is needed:
▪
A PC with MOM2 Win installed
See chapter 8 MOM2 Win
▪
A PC connected via Bluetooth
See section "Bluetooth set up"