STATUS heading on the front panel of the unit. This will tell the user
that the unit has detected a frame error.  

FAS Bit Errors

FAS bit errors are detected or inserted into the frame alignment signal
(FAS). They will appear every even frame from frame 0 to 15 in time slot
0.

E Bit Errors

The CRC MFAS is a 6 digit signal of 0 0 1 0 1 0 1 1 and is inserted bit-
by-bit into the first bit of the NFAS in frames 1, 3, 5, 7, 9 and 11. The
first bits of frame 13 and 15 are called E bits and are used to indicate the
data blocks with bit errors back to the transmitting side. If the E bit in
frame 13 = 0 it indicates that there is a CRC error in the data contained
in sub-multiframe I and if the E bit in frame 15 = 0 it indicates the same
situation for sub-multiframe II.

Bipolar Violation Errors & BPV Rate

The BPV error rate is calculated as follows:

BPV Error Rate = BPV Errors / # Total of Bits Received

This parameter is also a good measure for overall quality of service or
health of the E1 link. This is a measurement of the rate at which bipolar
violations are occurring over the entire E1 receive bit stream. 

Bit Errors & Bit Error Rate

The BERT-20E1 must be synchronized with a test pattern before it can
detect and record a bit error. Therefore, bit errors are registered only if a
BERT pattern is present. You can determine if the BERT-20E1 has
achieved pattern synchronization by observing the green PATTERN
SYNC LED under the RX STATUS heading on the front panel of the unit.
Each time the test set detects an error in the logic bit stream, one bit
error is recorded. If a bit error is detected, the unit will sound an

The display RESULTS arrow keys are
used to select a SUMRY (summary)
list of error results, NORM (normal)
standard list of BER results, or the
PERF (ITU performance) set of
results. The green LEDs to the left of
the LCD illuminate to indicate that
the corresponding set of results is
currently being displayed on the first
(two) lines of the LCD Display.

When the Display key is scrolled to
SUMRY, a list of BER results may be
viewed on the first (two) lines of the
LCD Display. Utilizing the RESULTS
up and down arrow keys the user
can scroll to the results that caused
the error. The red test summary
LED, located in the RX STATUS
section of the front panel,
illuminates when a bit error, BPV

error, FAS error, E-Bit error, CRC error, or E1 frequency deviation error is
detected.  

NORM – Normal List of BER and Non-BER Results
CRC Errors/CRC Error Rate

If the BERT-20E1 establishes frame synchronization with a PCM30C or
PCM31C framed signal, it will detect and record cyclic redundancy
Check (CRC) errors. CRC Errors are recorded if the E1 signal is live
traffic or a BER pattern. If a frame error is detected, the unit will sound
an audible alarm and blink the red test summary LED under the RX

VIEWING TEST RESULTS

BERT_20E1_6172-871_UG_V02  05/05/2004  8:24 am  Page 14

Содержание BERT-20E1

Страница 1: ...M BERT 20E1 2 048 Mbps Network Analyzer USER MANUAL BERT_20E1_6172 871_UG_V02 05 05 2004 8 24 am Page 1...

Страница 2: ...propriate surge and transient protection has been fitted The network and user terminals may be raised to hazardous voltages if this unit is connected to non SELV circuits This instrument should not be...

Страница 3: ...5 Software Revision Supplied Equipment Handling Precautions Front Panel Features Test Modes Frame Structures Test Patterns Line Codes Auxiliary Functions Power Batteries CONNECTING TO THE NETWORK 10 C...

Страница 4: ...cularly in environments where a convenient power supply may not be available Measurement Summary The BERT 20E1 combines BER test functions along with signal qualitymeasurement providing the ability to...

Страница 5: ...en the unit is initially powered up Supplied Equipment The BERT 20E1 is supplied with the following items A C power adapter User manual including CD ROM based E1 basics Carrying case Rubber boot Handl...

Страница 6: ...LUME controls BATTERY STAUS indicator TS CHANNEL SELECTION keys and display Stores test configuration Recalls test configuration Selects circuit termination Injects LOGIC BPV or FRAME CRC errors Power...

Страница 7: ...mode TONE Tone insert mode LLB Local loop back mode TLB Test loop back mode DLAY Round trip loop delay measurement CO64 Co directional 64kbps mode NONE No Framing P31 No multiframing P31C CRC multifr...

Страница 8: ...ITU T O 151 O 152 O 153 211 1 as per ITU T O 151 O 152 O 153 215 1 as per ITU T O 151 O 152 O 153 220 1 as per ITU T O 151 O 152 O 153 223 1 as per ITU T O 151 O 152 O 153 1 1 Alternating ones and ze...

Страница 9: ...BERT 20E1 batteries are not serviceable To recharge the batteries connect the 9 VDC power source to the DC jack located on the front panel The test set will charge during operation but it is recommen...

Страница 10: ...ction of 120 balanced data signals 120 TRANSMIT Transmitter output interface Provides 120 balanced data output EXT CLOCK IN Allows connection of a 75 timing reference as per ITU T G 811 The reference...

Страница 11: ...nuation Typically this gives a test signal 30 26 20dB down on the traffic signal Select MON monitor as your receive interface termination Next choose the appropriate MONITOR GAIN from the BERT 20E1 s...

Страница 12: ...4 Mode This mode allows the user to test in a fractional E1 environment The full 8 bits of data are available in each time slot The receiver and transmitter are manually configured for frame pattern a...

Страница 13: ...ing the logic error inject key the test set will insert a single logic error into the E1 payload At the same time a 1 millisecond counter is started and the receiver will start looking for the bit err...

Страница 14: ...E1 has achieved pattern synchronization by observing the green PATTERN SYNC LED under the RX STATUS heading on the front panel of the unit Each time the test set detects an error in the logic bit stre...

Страница 15: ...ed will be measured and the received level or amplitude will be displayed The BERT 20E1 demodulates the digitally encoded signal on the selected channel and drops the channel to the speaker The level...

Страница 16: ...rror seconds parameter is a one second interval of an available second where one or more bit errors have occurred The error free seconds parameter is a one second interval of an available second where...

Страница 17: ...to OFF The ERROR INJECT RATE menu can only be accessed when operating in the NORM AUTO F x 56 F x 64 D I LLB or TLB modes 03 User Pattern 1 The user pattern 1 is a user defined patterns that is 32 bi...

Страница 18: ...l Fx56 Fx64 LLB or TLB and the printer is correctly hooked up and configured correctly The following sample printout will be observed TEST RESULTS TIME CURRNET TIME DATE CURRENT DATE MODE NORM FRAME P...

Страница 19: ...ular DIN Custom EXT CLOCK IN 75 BNC Auxiliary Power DC power jack accepts 2 5 mm diameter female plug with center or tip and outside Receiver Modes NORM Tx follows Rx AUTO Auto Configure Rx Tx is inde...

Страница 20: ...G 706 G 823 G 732 and I 431 ETS 300 011 300 233 and 300 166 CTR12 and CTR4 Transmitter Line Code AMI or HDB3 Framing NONE PCM30 PCM30C PCM31 and PCM31C Frequency 2048000 Hz 2 5 ppm Line Coding AMI HDB...

Страница 21: ...ORM SUMMARY PERFORMANCE Clock Slips A positive number indicates that the reference clock is faster a negative number indicates that the reference clock is slower Errors Logic rrrors logic error rate B...

Страница 22: ...rsonnel The test set circuit contains static sensitive devices and care must be taken in handling the printed circuit board If the protection of a test set has been impaired it should not be used and...

Страница 23: ...23 BERT_20E1_6172 871_UG_V02 05 05 2004 8 24 am Page 23...

Страница 24: ...30 16 08 90 F 33 0 1 34 61 23 77 OTHER TECHNICAL SALES OFFICES Toronto CANADA Sydney AUSTRALIA Mumbai INDIA Madrid SPAIN and the Kingdom of BAHRAIN Megger products are distributed in 146 countries wo...

Отзывы: