AVO®830 and AVO®835
7
Safety
Safety
Safety Information
Retain these safety warnings for future reference.
Read and understand the safety information before using the instrument.
The following warnings must be obeyed.
Important: This instrument must be used only by suitably trained and competent persons.
Users of this instrument and their employers must do a valid risk assessment of all electrical work to identify
potential sources of electrical danger and risk of electrical injury, and must conform to national safety legislation.
Where assessment shows significant risk from high energy systems, the use of fused leads must be considered.
If the instrument is used in a manner not specified by the manufacturer, the protection provided by the
instrument may be impaired.
The instrument must not be used if any part of it is damaged.
Use only Megger approved test leads and accessories with this instrument.
Test leads and probes must be in good order, clean and have no broken or cracked insulation.
Hold the test probes behind the finger-guards when they are in contact.
Always check that the correct instrument connections and range selection have been made before
connection to the test subject.
Voltage or current that exceeds the instrument ratings can damage the instrument and present a shock or fire
hazard.
Use caution when working with voltages above 30 VAC rms, 42 VAC peak or 60 VDC. These voltages present
a shock hazard.
To avoid the risk of electric shock do not connect the thermocouple to circuits greater than 10 V.
Before any resistance, continuity, diode, or capacitance measurements are started, circuits must be de-
energised and isolated from the mains power.
Before and after use always test a known voltage to check for correct instrument operation. Do not use the
instrument if incorrect results are obtained.
Make sure that the test probes are disconnected before the case or battery cover is opened.
Replacement fuses must be of the correct type and rating.
The instrument must not be used in wet conditions. If it should become wet, it must be dried thoroughly
before use.
Test leads, which are used for mains measurements, must be rated as appropriate for measurement category
IV or III. They must also have a voltage rating of at least the voltage of the circuit to be measured.
Содержание AVO 830
Страница 1: ...AVO 830 and AVO 835 True RMS Digital Multi Meter User Guide ...
Страница 2: ......
Страница 44: ...www megger com AVO 830 and AVO 835 44 ...
Страница 45: ...www megger com AVO 830 and AVO 835 45 ...
Страница 46: ...www megger com AVO 830 and AVO 835 46 ...
Страница 47: ...www megger com AVO 830 and AVO 835 47 ...