www.megger.com
AVO®830 and AVO®835
30
Measurements
AC Voltage Detection
To detect AC voltage, the top of the instrument must be placed close to a live conductor.
The instrument gives an audible, as well as a visual, indication when voltage is detected.
Caution:
If there is no voltage indication, voltage could still be present. Do not rely on this detection
when used in the vicinity of shielded wire. Operation may be effected by differences in socket design,
insulation thickness and type.
Hz
CAT III 1000V
CAT IV 600V
COM
FUSED
100mA
V
10A
Ω
AVO835
RANGE
MODE
MIN MAX
AVG
REL %
REL
Low Z
HOLD/Auto
ac+dc
APO
Lo
Hz
Hz
°F
Hz
V
Hi
A
mA
OFF
OFF
Ω
°C
MODE
Lo
Hi
HI
LO
Fig 11: AC voltage Detection Set-up
There are two sensitivity settings:
To Check the Presence of AC Voltage
1. Set-up the instrument as shown in Fig 11.
•
Set the Function Dial to
.
2. Place the instrument close to a live conductor.
3. Press
MODE
to select:
•
HI
: (default) Highest sensitivity to AC voltage
•
LO
: Low sensitivity to AC voltage