background image

27

Pressure Adjustments and Tests

1. Make sure the power source is 120 Volt 60 Hz. Use a power regulator 

 

to stabilize the power voltage in order to increase accuracy of pressure setting.

2. Attach the test kit to pump and manometer. 
3. Press the Power Micro Switch.
4. Perform the pressure test procedure from Minimum Pressure (LEG) 

 

to Maximum Pressure (FT).

Minimum Pressure Test

1. Press Mode Membrane Button (CALF/THIGH SLEEVES) and LED Situation 
Monitor will display “LEG”.
2. Check the pressure is 40 mmHg ( +10/-5 mmHg ).

Maximum Pressure Test

1. Press Mode Membrane Button (FOOT) and LED Situation Monitor will display ”Ft”
2. Check the pressure is 120 mmHg ( +10/-5 mmHg ).

Pressure Adjustment Procedure

1. Open the case and identify pressure regulators.

 

2. Pressure adjustment can be adjusted with a Philips screwdriver.  

Pressure can be adjusted by clockwise/anti-clockwise turns to 

 

decrease/increase the pressure respectively. 

   

Содержание Hemo-Force MDS600INT

Страница 1: ...1...

Страница 2: ...5 Control Panel Features 5 Operating Instructions 5 Alarms and Indicators 8 Cleaning 10 Electromagnetic Compatibility 11 Technical Specifications 14 Servicing and Maintenance 16 General Disassembly an...

Страница 3: ...e the state corresponding to a specified initial number Battery indicator that can show remaining battery status Unique hanging bracket that provides easy attachment Intended Use The system consists o...

Страница 4: ...ith hospital policy 5 Keep the pump and sleeve away from sources of liquid and open flames 6 The control unit should only be repaired by an authorized professional 7 Do not drop the control unit or st...

Страница 5: ...Flammability as defined in CPSC 16 CFR Part 1610 Control Panel Features Alarm Light Blue LED Display Window Battery Indicator Time in Hours Foot Mode Pressure Settings for Foot at 120mmHg Time Reset...

Страница 6: ...two finger rule to check that the correct fit is used Insert 2 fingers horizontally with palm up and both fingers touching the patient s leg in between the sleeve and the patient This practice should...

Страница 7: ...d hold the power button to turn pump on when air hose s and sleeve s are properly connected 9 Pump will start initialization process lasting approximately 10 seconds 10 The pump will automatically det...

Страница 8: ...may negatively impact the effectiveness of the pump and or cause a skin integrity issue for the patient Make sure that there are no kinks in the tubing and that the connectors are properly locked Neve...

Страница 9: ...air hose sleeve connectors Attach sleeves to pump and restart Ensure sleeves are fully deflated before restart Check sleeve configuration and restart Check connectors for any blockage and restart Refe...

Страница 10: ...ly before cleaning and inspection The Sleeve It is a single patient use sleeve Do not wash or clean discard after use Caution Do not use phenol based cleaning solutions Switch off the electrical suppl...

Страница 11: ...ce d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitte...

Страница 12: ...1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power qual ity should be that of a typical commercial or hospital environment interruptions and voltage vari...

Страница 13: ...ransmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the fol...

Страница 14: ...s Inflation Time 8 Seconds Hold Time 4 Seconds Deflation Time 18 Seconds Display Type LED Alarm Type Audio and Visible Hook attachment Yes Owners Manual Included Battery Pack Model optional MDS600BAT...

Страница 15: ...15 Symbols Used Type BF Applied Part Manufacture Date of Manufacture Batch Code Ex YYMMDD01XX Serial Number Caution...

Страница 16: ...ain Relief Bushing 6 Control Panel 6 1 Power Micro Switch ON OFF 6 2 Mode Membrane But ton Foot 6 3 Membrane Button Reset for timer reset 7 Function Status LED Light Indication 7 1 Power Light Green 7...

Страница 17: ...ery Pack recharge discharge 16hr 4hr Optional equipment C Initial Assessment Procedure Item Tasks 1 Check the Plastic Chassis Housing appearance 2 Check the Mount Fasteners 3 Check the Hooks 4 Check t...

Страница 18: ...run for a minimum of 1 hour and check calibration Adjust if required Operational Maintenance Routine checks are recommended to be carried out for each time the system is in use For the control unit 1...

Страница 19: ...ter 1 Identify the air filter module on the back the device 2 Press the side lock to release the cap 3 Clean the cap with a soft brush 4 Remove the filter and clean the bottom with a soft brush 5 Inse...

Страница 20: ...e back of the device 2 Unscrew the screw to release the cable cover 3 Press the side lock to release the battery pack 4 Remove the battery pack and check the socket for dirt and or debris 5 Inspect th...

Страница 21: ...Pack Placement procedure 1 4 to remove the battery pack 3 Unscrew the 6 pan head screws 4 Open the case and keep the screws in a safe place 5 Disconnect 2 air tubes from the valve 6 The control unit...

Страница 22: ...sor and valve 4 Unscrew and remove the 2 screws securing the compressor 5 Remove the whole compressor set out of the case 6 Re assemble new components in reverse order Solenoid valve set replacement 1...

Страница 23: ...23 2 Unplug the valve power cable from PCB board 3 Disconnect the air tube from the valve 4 Unscrew and remove the 2 screws 5 Remove the whole valve set 6 Re assemble new components in reverse order...

Страница 24: ...24 Fuse replacement 1 Identify the fuse module 2 Remove the fuse s 3 Insert the new fuse s 22 23 17 16 21 19 18 20 Schematics...

Страница 25: ...PCB Set 10 MDSCH40020 120V Power Cord 11 MDSCML20045 Tube Set For Compressor 12 MDSCML20046 Pressure Tube Set 13 MDSCML20047 Air Tube Set 14 MDSBF004001 Rubber Foot 15 MDSAL200208 Label Front Panel Fo...

Страница 26: ...ssure at 120 mmHg Exclusive to Hemo Force IPCS Battery Indicator Time in Hours Time Reset Button Power Button Power Light Single Sleeve Mode Calf Thigh Modede Foot Mode Alarm Light Pressure Display Wi...

Страница 27: ...Minimum Pressure Test 1 Press Mode Membrane Button CALF THIGH SLEEVES and LED Situation Monitor will display LEG 2 Check the pressure is 40 mmHg 10 5 mmHg Maximum Pressure Test 1 Press Mode Membrane B...

Страница 28: ...obligation under this warranty is limited to the repair of equipment returned to the place of purchase within 12 months of delivery date We agree to service adjust any equipment returned and to replac...

Отзывы: