Opening the case/interal inspection
1. Place the device on a soft surface with the front case facing down.
2. Perform “Battery Pack Placement” procedure 1-4 to remove the battery pack.
3. Unscrew the 6 pan head screws.
4. Open the case and keep the screws in a safe place.
5. Disconnect 2 air tubes from the valve.
6. The control unit is now ready for internal inspection
Compressor replacement
1. Identify the compressor internally.
2. Unplug the power connector.