47
Table 6.4
Alarm and Alert Troubleshooting
Alarm/
Alert
Message Displayed
Alarm/Alert
Description
System Status
Troubleshooting
17
Patient Temperature
calibration error.
Restart system.
Alarm
In Automatic Mode,
internal temperature
calibration from
Patient Temperature
1 probe input is not
functioning.
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Turn the Control Module off. Wait 30 seconds and
turn power back on. If the error persists, replace
the Isolation Card and retest. If the error persists,
call Medivance Service. Calibrate upon
completion.
18 Patient
Temperature
calibration error.
Restart system.
Alert
In Manual Mode,
internal temperature
calibration from the
Patient Temperature
1 probe input is not
functioning.
Current Mode
continues. Audio
alarm half second
tone every 2
minutes. Alarm/Alert
indicator flashing.
Turn the Control Module off. Wait 30 seconds and
turn power back on. If the error persists, replace
the Isolation Card and retest. If the error persists,
call Medivance Service. Calibrate upon
completion.
19 Patient
Temperature
calibration error.
Restart system.
Alarm
In Automatic Mode,
internal temperature
calibration from the
Patient Temperature
1 probe input is not
functioning.
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Turn the Control Module off. Wait 30 seconds and
turn power back on. If the error persists, replace
the Isolation Card and retest. If the error persists,
call Medivance Service. Calibrate upon
completion.
20 Patient
Temperature
calibration error.
Restart system.
Alert
In Manual Mode,
internal temperature
calibration from the
Patient Temperature
1 probe input is not
functioning.
Current Mode
continues. Audio
alarm half second
tone every 2
minutes. Alarm/Alert
indicator flashing.
Turn the Control Module off. Wait 30 seconds and
turn power back on. If the error persists, replace
the Isolation Card and retest. If the error persists,
call Medivance Service. Calibrate upon
completion.
21 Patient
Temperature
2 high. Check
patient temperature.
Alarm
Patient Temperature
2 reading is above
38.
◦
C (100.4°F) and
water temperature is
above 38.
◦
C
(100.4°F) and
system is unable to
cool in Automatic
Mode
Changes to Stop
Mode. Audio alarm
half second tone
every 2 minutes.
Alarm indicator
flashing
Operate the system in Manual Mode with a setpoint
below 38°C until the water temperature is less than
38°C. Press Automatic and retest. If the error
persists, see General Troubleshooting - Will Not
Cool.
22 Patient
Temperature
2 above user defined
limit.
Alert
Patient Temperature
2 reading is equal or
greater than the
user-defined value
established in the
custom settings
menu.
Current Mode
Continues. Audio
alarm half second
tone every 2
minutes. Alarm/Alert
indicator flashing
Adjust Patient High Temperature Alert in the
custom menu if desired.