79
APPENDIX A SPECIFICATIONS
Parameter Specification
Control Modes
Automatic, Manual, Purge, Stop
Heater Capability
750 W
Circulating Fluid
Distilled or Sterile Water
Reservoir Capacity
5 liters
Water flow rate (total)
0.5 - 8.0 liter/min
Patient Probe Type
YSI 400 Series compatible
Number of Patient Probes
2 (1 monitor, controls, 1 monitor,
alarms)
Patient Temperature Display Range
Resolution
10°C to 42°C, 50°F to 107.6°F
in 0.1°C/
°
F
Patient Temperature Measurement Accuracy ±0.4ºC (10ºC to 32ºC)
±0.2ºC (32ºC to 38ºC)
±0.4ºC (38ºC to 42ºC)
(includes
±
0.1ºC external probe)
Patient Temperature Control Range -
Automatic Mode
33°C to 37°C, 91.4°F to 98.6°F
in 0.1°C/F increments
Water Temperature Display Range
3°C to 45°C, 37.4°F to 113.0°F
in 1°C/F increments
Water Temperature Control Range -
Manual Mode
4°C to 42°C,39.2°F to 107.6°F
in 0.1°C/F increments
Maximum Water Temperature
(Automatic Mode)
32°C to 42°C, 89.6°F to 107.6°F
in 1°C/F increments
Minimum Water Temperature
(Automatic Mode)
4°C to 32°C, 39.2°F to 89.6°F
in 1°C/F increments
Mains Input*
115VAC, 60 Hz, 11 A (nominal)
230VAC, 50 Hz, 5.5 A (nominal)
Current Leakage
< 300uA
Circuit Breaker
12 Amp
*Mains supply voltage fluctuations are not to exceed
±
10% of nominal supply voltage.