29
Table 5.1
Model 2000 Verification Data
Test Criteria
Flow Rate
Factor H - 300ml/min <
Flow < Factor H +
300ml/min
Inlet water pressure
-7.2 psi < P < -6.8 psi
Patient Temperature 1 PT1 (4 checks)
-0.3
°
C < CK1 < 0.3
°
C
-0.3
°
C < CK2 < 0.3
°
C
-0.1
°
C < CK3 < 0.1
°
C
-0.3
°
C < CK4 < 0.3
°
C
Patient Temperature 2 PT2 (4 checks)
-0.3
°
C < CK1 < 0.3
°
C
-0.3
°
C < CK2 < 0.3
°
C
-0.1
°
C < CK3 < 0.1
°
C
-0.3
°
C < CK4 < 0.3
°
C
Patient Temperature Echo Out (4 checks)
-0.2
°
C < CK1 < 0.2
°
C
-0.2
°
C < CK2 < 0.2
°
C
-0.2
°
C < CK3 < 0.2
°
C
-0.2
°
C < CK4 < 0.2
°
C
Water Temperature T1 (3 checks)
-0.4
°
C < CK1 < 0.4
°
C
-0.4
°
C < CK2 < 0.4
°
C
-0.2
°
C < CK3 < 0.2
°
C
Water Temperature T2 (3 checks)
-0.4
°
C < CK1 < 0.4
°
C
-0.4
°
C < CK2 < 0.4
°
C
-0.2
°
C < CK3 < 0.2
°
C
Water Temperature T3 (3 checks)
-0.4
°
C < CK1 < 0.4
°
C
-0.4
°
C < CK2 < 0.4
°
C
-0.4
°
C < CK3 < 0.4
°
C
Water Temperature T4 (3 checks)
-0.4
°
C < CK1 < 0.4
°
C
-1.0
°
C < CK2 < 1.0
°
C
-1.0
°
C < CK3 < 1.0
°
C
Miscellaneous
Heat
Cool
Flow
Heat < 120 seconds
Cool < 150 seconds
Factor H - 300ml/min <
Flow < Factor H +
300ml/min
Secondary Alarm
Alarm 37 in < 18 seconds