7.1.2 Treadmill belt set to the correct tension
Before initiating the system check if the belt is in the correct tension. If it is too loose, perform
tightening using the two hexagonal screws that are adjusting the back pulley.
7.1.3 Perturbation plate centering
After switching the device ON confirm the perturbation platform is centered to the stationary part of
the device. If not, alert for maintenancy.
7.2 Immediate maintenance
Immediate maintenance is necessary if:
The device has been under high mechanical stress (push, power supply cable and/or interface
cable defect through driving over it or pulling it)
Lateral moving platform is stuck (from some obstacle or from any non-visible cause)
Fluid has got into the device
Cable and/or connector plug have been damaged
Coverings and/or safety warnings have fallen Off
The rope or the deflection pulleys show sign of wear
A defect or malfunction of the device has been detected or is suspected
The BalanceTutor software crashes and/or restarting
Strange periodical mechanical noises come from motor systems
Only a properly and regularly serviced device is safe. The maintenance of the devices has to be
performed by the service engineers authorized by MediTouch, preferably within the scope of a
maintenance contract.
7.3 Regular inspections/examinations
Following list will containe which elements to be examined or inspected every year by MediTouch
or any any technical personel authorized by MediTouch:
Check metal construction for: Visual damage, rust, paint scraped, metal deformation.
Confirm the BalanceTutor software running with the latest version.
Running belt tension adjusted to the correct tension
No backlash on lateral transmission system
Electrical safety test; Protective Earth Resistance, Isolation Resistance and Leakage current.