including the outlets and an authorized electrical engineer must check these points for perfect
functions regularly (1 ... 4 years). Inspections of the electrical installations within the
building are not incumbent on the supplier MediTouch.
As standard, a dedicated power line of 230 Volt AC, 50/60 Hz is used with a 16A C type
circuit breaker. In case of countries like USA where the voltage of a two phase in a Y
constellation is 208V, you need to use a the following wiring L1/L2/PE 50/60Hz 2x15A
circuit breaker with a dedicated power line.
BalanceTutor supplied with an external isolation transformer it required to be fixed to a wall
and not be disconnected in any case.
Before installing the BalanceTutor system please compare the specifications on the
nameplate concerning the mains voltage and the mains frequency with your local
characteristics. Connection only if identical.
Check the main lead, the voltage power supply outlet and ground wire protection-contacts
before plugging it in.
Damaged leads and couplers and defective or dirty contacts have to be exchanged
immediately. Rubber-leads can get porous and friable after some years.
For further questions please ask your electrical engineer or MediTouch.
Plug the device directly into the wall socket with checked ground wire. Device has
to be connected to a separate circuit. The use of an extension cables or a multiple
plug sockets is not allowed.
Electrical devices with mains connections must neither be used in wet and humid
areas (e.g. swimming pool, sauna, etc.) nor in environmental chambers.