background image

ASM

Medicare

SUNDOWN MkII MATTRESS ELEVATOR

USER INSRUCTIONS

CODE 09760A

With Emergency Backup Battery

& Safety Anti Trap Actuator.

Better By Design.

Picow Farm Road

Runcorn, Cheshire

Tel: 01928 574301

Fax: 01925 575130

Email:- 

[email protected]

www.asmmedicare.com

Sundown MkII

Mattress Elevator

User Instructions

March-03

Содержание SUNDOWN MkII

Страница 1: ...NS CODE 09760A With Emergency Backup Battery Safety Anti Trap Actuator Better By Design Picow Farm Road Runcorn Cheshire Tel 01928 574301 Fax 01925 575130 Email sales asmmedicare com www asmmedicare com Sundown MkII Mattress Elevator User Instructions March 03 ...

Страница 2: ...lation 4 Electrical Fittings 5 Care Maintenance 6 Trouble Shooting 7 Spare Parts List 8 Guarantee 9 Technical Data Blue Walker Trolley 02138 Corner Shower Stool 02010 Solent Overbed Table 6020 Wren Perching Stool 02303HB ASM Medicare ASM Medicare ...

Страница 3: ...e above are not included please contact your supplier immediately stating the serial number date of purchase Please Note A The emergency battery provided will allow the elevator to be repositioned in the event of power failure This battery should be replaced every 6 months or after one use B The fitting straps may obstruct access to divan drawers Always replace securely if removed for access Thank...

Страница 4: ...at some type of beds I e metal framed may not be suitable for this product 9 Technical Data Product Weight 16 05kg 35 3lb MUW 190kg 30st Overall Dimensions Closed Width 941mm Depth 710mm Height 163mm Overall Dimensions Open Width 1566mm Depth 710mm Height 655mm Lift 10 Minimum 80 Maximum Power Supply 240volt 50Htz To Actuator Power System 24volt DC Via Integral Control Box Emergency Back Up Batter...

Страница 5: ...nect the retaining and locking buckles Fig 2 DO NOT cross the two long straps e Tighten by pulling the straps through the locking buckles Once tight tie off for extra security f Pass the long fitting strap in the centre of the elevator across the length of the bed through under the bed base Pass through the rear bracket at the centre of the elevator fit the locking buckle Fig 3 in the centre of th...

Страница 6: ... connection secure or lifting loose if necessary Emergency backup Backup battery button not Push backup battery button battery not functioning being depressed in unison simultaniousley with handset with handset Elevator unit moves Fitting straps loose Retighten all straps on when operated bed base Mattress moves when Fitting strap loose Retighten mattress elevator operated fitting strap ASM Medica...

Страница 7: ...mply Push Together And Click Firmly Into Place Fig 3 Fig 4 6 Spare Parts List LAC 1 Actuator with integural Control Box LA 2 Actuator Only LHF 4 Handset LMP 5 Power Lead LSFS 6 Fitting Straps x 4 LSPS 7 Short Width Fitting Straps with Buckles x 2 LLPS 8 Short Length Fitting Strap with Buckle AMS 9 Mattress Fitting Strap with Buckle ASM Medicare ASM Medicare Fig 2 ...

Страница 8: ...n on the actuator and simultaneously depressing the desired button on the handset WARNING Keep limbs away from ALL moving parts rear stops Should entrapment occur stop operation of the elevator seek medical advise for injury if sustained 5 Care Maintenance Your new Adjustable Mattress Elevator should only be serviced by a trained competent engineer Always change the backup battery every six months...