Appendix B EMC Guidance and Manufacturer’s Declaration
Basic User Manual
195
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) tele-
phones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the EQUIPMENT is used exceeds the applicable RF compliance
level above, the EQUIPMENT should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating
the EQUIPMENT.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.