6 Acquiring Images
Basic User Manual
95
1.
Enter the inactivated B+M mode, tap
Steer M
on the touch screen to set the number
of M lines.
Multiple M-lines display after selecting the desired number as the following figure
shows.
Figure 6-12 Inactivated Anatomical M-Mode Imaging Screen
2.
Adjust the position and the angle of M line.
−
Press the confirm key on the control panel to select the desired M line.
−
Position the desired M-line by using the trackball.
−
Rotate the
Angle
knob on the control panel to adjust the angle of M line.
3.
Press the
Update
key on the control panel to activate the anatomical M-mode.
NOTE:
Once the anatomical M-mode is activated, you can move the trackball to stop the
M trace and adjust the M line. The system continues tracing if the trackball is idle
for more than 0.5s.
4.
Optimize anatomical M-mode images. For details, refer to Section 6.4.3 Optimizing
M-Mode Images.
5.
Press the
M
key again to exit the screen.
6
�
4
�
3 Optimizing M-Mode Images
The touch screen displays all controls related to the M-mode imaging (as shown in the
figure), you can use them to optimize M-mode images.
10
2D Imaging
5
M-line
0
Содержание P25 EXP
Страница 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Страница 5: ......
Страница 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Страница 48: ...This page is intentionally left blank...
Страница 87: ...75 6 Acquiring Images You can acquire images by optimizing the relevant parameters for clinical diagnosis...
Страница 168: ...This page is intentionally left blank...
Страница 176: ...This page is intentionally left blank...
Страница 200: ...This page is intentionally left blank...
Страница 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...