9 3D Imaging
134
Basic User Manual
NOTE:
This setting is only effective when entering 3D imaging from a frozen B mode image.
9
�
2
�
11 Adjusting ZAngle/ZScale
The image may be distorted if you perform the scan too quickly by using the probe. To
acquire a vivid image, you can adjust the value of
ZAngle
/
ZScale
before the scan.
●
If
Scan Mode
is set to
Sector
, you can adjust the sweep angle when performing the
scan by the probe. Tapping the left/right part of
ZAngle
can adjust the angle.
●
If
Scan Mode
is set to
Linear
, you can adjust the sweep span when performing the
scan by the probe. Tapping the left/right part of
ZScale
can adjust the scale.
NOTE:
This setting is only effective when entering 3D imaging from a frozen B mode image.
9
�
2
�
12 Restoring the Image
Tap
3D/4D
>
Reset
to restore the relevant settings.
9
�
2
�
13 Customizing Presets
You can modify parameters, adjust sample volume or ROI in the 3D mode and save them
as the application preset for future use. Moreover, the customized layout of the application
mode screen can be modified and restored to factory defaults.
NOTE:
Only the volume probes are capable of this feature.
For details, refer to Section 6.1 Selecting a Probe and an Exam Type.
The application presets can be displayed as options of the
User Mode
item in the
inactivated 3D
mode.
9
�
2
�
14 Performing Follicle Auto-Measurement
Tap
AVC Follicle
>
Left Ovary
/
Right Ovary
to automactically measure the number and
the volume
of follicles, and the position of follicles is displayed on three reference images and 3D
image.
Содержание P25 EXP
Страница 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Страница 5: ......
Страница 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Страница 48: ...This page is intentionally left blank...
Страница 87: ...75 6 Acquiring Images You can acquire images by optimizing the relevant parameters for clinical diagnosis...
Страница 168: ...This page is intentionally left blank...
Страница 176: ...This page is intentionally left blank...
Страница 200: ...This page is intentionally left blank...
Страница 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...