115
Deutsch
Electrostatic
transient/
burst
IEC 61000-4-4
± 2 kV for power
supply lines100
kHz repetition
frequency ±1 kV
for input/output
lines
± 2 kV for power
supply lines100
kHz repetition
frequency ±1 kV
for input/output
lines
Mains power quality should
be that of a typical commer
-
cial or hospital environment.
Surge IEC
61000-4-5
±0.5 kV, ±1kV
differential mode
line-line
±0.5 kV, ±1kV
differential mode
line-line
Mains power quality should
be that of a typical commer
-
cial or hospital environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
0% UT (100% dip
in UT ) for 0.5 cy
-
cle at 0°, 45°, 90°,
135°,180°, 225°,
270°, and 315°
0% UT (100%
dip in UT) for 1
cycleat 0°
70% UT (30% dip
in UT ) for 25/30
cycles a t 0°
0% UT(100% dip
in UT ) for 250/
300 cycle at 0°
0% UT (100% dip
in UT)for 0.5 cycle
at 0°, 45°, 90°,
135°,180°, 225°,
270°, and 315°
0% UT (100% dip
in UT) for 1 cycle
at 0°
70% UT (30% dip
in UT) for 25/30
cycles at 0°
0% UT(100% dip
in UT) for 250/
300 cycle at 0°
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user
of the Model WDP-303936
product name requires
continued operation during
power mains interruptions,
it is recommended that
the Model WDP-303936 be
powered from an uninter
-
ruptible power supply or a
battery.
Power frequen
-
cy (50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m,50/60H
30 A/m,50/60Hz
Power frequency magnetic
fields should be at levels
characteristic of a typical lo
-
cation in a typical commer
-
cial or hospital environment.
NOTE: UT is the a. c. mains voltage prior to application of the test level.