19
Instructions for Use
Loading a Patient Cassette for double pump mode
10. Sterile operator: Connect the end of the tubing marked OUTFLOW CANNULA tubing to the cannula, and the tubing
marked SHAVER to the suction port of the shaver hand piece. If a shaver is not used, the tubing can remain
disconnected.
Connect the end of the tubing marked RF PROBE to the RF Probe. If a Double Pump RF Patient Cassette is used but
a RF Generator is not used, this tubing can remain disconnected.
Priming the Day Cassette and Irrigation Tubing
To optionally slowly prime the Day Cassette and Irrigation Tubing with saline, press the Soft Key ‘Prime’ on the Run/
Stop menu. The pump will automatically prime the tubing, counting down from 10 sec.
To stop the priming, press the Soft Key . When priming stops, the Display will return to the Run/Stop menu. Priming
can be repeated if necessary. Pressing Run/Stop during Priming, the pump starts Running immediately.
Adjusting the pressure and outflow
The Run/Stop menu displays the pressure and outflow settings for the selected joint type. The initial settings can be
adjusted using the Pressure increase/decrease buttons and the Flow increase/decrease buttons. Flow adjustment is only
relevant if a Patient Cassette and an outflow cannula is being used (double pump mode) as the Flow settings refer to the
outflow from the joint as controlled with the Patient Cassette. Pressure can be set to 20-180 mmHg. Outflow can be set to
0-600 ml/min, but will be reduced at high pressures.
Starting the fluid flow
To start the fluid flow, press the Run/Stop button on the front panel. The pump will run and control the pressure and
fluid outflow. Fluid outflow is only controlled if a Patient Cassette or Patient Cassette RF is used.
Stopping the fluid flow
To stop the fluid flow, press the Run/Stop button on the keyboard. When the flow is stopped, pressing the Soft Key will
take you back to the procedure selection menu.
Removing the Irrigation Tubing when the Day Cassette will be used again the same day
1. Stop the pump and close all clamps on the Day Cassette and the red clamp on the Irrigation Tube.
2. Disconnect the Irrigation Tubing near the check valve, leaving the used check valve attached to the Day Cassette.
The used check valve must stay to maintain sterility of the connection on the Day Cassette.
3. Discard the disconnected Irrigation Tubing in accordance with local hospital procedures.