14
Unpacking, Assembly and System Check
Verify that all items have been received and are not damaged. Damage should be reported immediately to the carrier and to
your Medical Vision AB representative.
PLEASE NOTE:
Save all containers and packaging material. They will be required if the Double
Pump RF is shipped.
Unpacking
Assembly and System Check
1. Fit the front visor by holding it with
both hands as illustrated. Note the
angle of the arrows and slide both
sides in gently as illustrated in the
second picture. Thereafter turn the
vi sor up fully and push both sides
simultaneously all the way in. Close
the front visor with both hands.
See illustrations on pages 11 and 12.
2. Connect the mains cable to the mains receptacle
(16) on the rear panel of the Pump. Connect the other end of the
mains cable to a grounded electrical outlet.
3. Attach the Pump Foot Control Receptacle
(11) on the front panel.
4. If the pump shall be interconnected to a shaver and/or a RF generator to form a system, ensure that this is done to
achieve optimal and practical use, for instance by placing all devices the same rack/tower, to avoid long cable routing.
5. If an ArthroCare
®
, Quantum generator is used, plug the RF Cable A1 into the RF Cable Receptacle
(17) on the rear of
the Pump and into the Foot Control Receptacle of the RF Generator. Connect the RF Generator Foot Control into the RF
Generator Foot Control Receptacle (12) on the front of the Pump. See Connection diagrams on pages 4 and 5. See
Connection diagrams on pages 6 and 7 for other RF systems.
6. There are several alternatives for connecting a shaver system to Double Pump RF as shown in the Connection dia-
grams on pages 4 -- 7.
7. If the shaver is to be controlled by the Double Pump RF an appropriate Shaver Interface Cable must be used. Refer
to page 30 of this manual to select the right Shaver Interface Cable. Connect the Shaver Interface Cable
(6) to the
Shaver Interface Cable receptacle
(19) on the rear panel of the Pump, and connect the other end to the foot control
connector of the shaver console
(4). A 4 pedal Foot Control REF No. 1404176 must be used.
8. If the shaver, having a finger controlled hand piece, is to control the Double Pump RF an appropriate Shaver
Interface Box must be used. Refer to page 30 of this manual. Connect the cable of the Shaver Interface Box to
the Shaver Interface Box Receptacle
(20) on the rear panel of the Pump. The Shaver Interface Box is fitted
on the front panel of the shaver console
(4), and the shaver hand piece is connected to the box as shown in the
Connection diagram. Attach the Pump Foot Control to the Pump Foot Control Receptacle
(11) on the front panel.
A 2 or 4 pedal Foot Control REF No. 1404003 or 1404176 respectively must be used.
9. Turn the On/Off Switch on the rear panel of the pump to the ‘On’ position. The system will proceed through a self test.
Turn on the pump by pressing On/Standby button (2) and the software version number will be displayed. During this
sequence, verify that the Display and LED lights are working properly and that the alarm tone is audible. The Display
should show ‘START’ and ‘Setup’ options associated with the Soft Keys
(5) or procedure selection menu if Quick Start
is enabled. If an error code appears instead, the pump should not be used. Contact Medical Vision AB representative,
and report the error code number.
10. To change display text language: Select “Setup” by pressing the right lower Soft Key (5). In the SETUP menu , scroll
down with the left lower Soft Key (5) until the display shows “Language”. Press OK, and scroll down with until the
display shows the users’ language. Press OK, and escape from the Setup menu by pressing twice.
11. If Double Pump RF malfunctions or shows signs of damage, contact Medical Vision AB representative.