8
Principle of Operation
The Double Pump RF can operate in two different modes:
1)
Single (Inflow only)
2)
Double pump
During
Single mode, the Double Pump RF regulates the pressure using the Double Pump RF Day Cassette only.
During
double pump mode, the Double Pump RF controls the pressure and outflow using the Day and Patient Cassettes.
For optimal performance in double pump mode, an Outflow Cannula is recommended. The Double Pump RF Day Cassette is
connected to the sterile saline bags and the Irrigation Tubing. The Irrigation Tubing is connected to the arthroscope sheath.
The Irrigation Tubing is delivered separately or in the Double Pump RF Patient Cassette package.
The Double Pump RF Patient Cassette discharges the liquid from the joint into a waste container.
The Double Pump RF has the SmartVision
®
feature. SmartVision
®
will detect blood and/or debris when it emerges from the joint.
It will automatically elevate pressure into the joint and outflow when blood is detected, or elevate outflow from the joint when
debris is detected. SmartVision
®
can be disabled if desired.
The pump also has an Outflow Tracking feature, which controls the inflow to the joint to replace the amount of liquid removed
by the outflow pump. The result is a constant volume of fluid in the distended joint. This helps to prevent collapse of the joint in
cases with highly compliant joint. Outflow Tracking can be enabled if desired.
Furthermore, the system automatically compensates for liquid resistance in the irrigation system. This means that the pressure
value displayed is the actual pressure in the joint, not the pressure at the pump or in the Day Cassette.