25
System Care
System Environmental Requirements
Equipment Disposal
Surface Cleaning of Double Pump RF, Foot Controls and system devices
The Double Pump RF may be safely stored and transported at an ambient temperature range of -40°C (-40°F) to +70°C
(+158°F), relative humidity between 10 and 85%, and atmospheric pressure of 500 hPa (7.2 psi) to 1060 hPa (15.3 psi).
The Double Pump RF may be safely operated at an ambient temperature range of 10°C (65°F) to 40°C (104°F), relative
humidity between 10 and 85%, and atmospheric pressure of 700 hPa (10.1 psi) to 1060 hPa (15.3 psi).
PRECAUTION
: Storing or operating Double Pump RF outside these temperature ranges may cause the Pump to malfunction
or to cause damage to the Pump.
Double Pump RF contains electronic printed circuit board assemblies and should not be disposed in any waste container.
It should be disposed of in accordance with any applicable national or institutional policies relating to obsolete electronic
equipment. Contact your Medical Vision AB representative for return of Double Pump RF for proper disposal. Discard the
disposables in accordance with local hospital procedures for potentially contaminated items. Double Pump RF does not
contain NiCd or lead batteries.
Wipe the pump and foot controls with a soft damp cloth. Alcohol based agents may be used.
PRECAUTION:
DO NOT
apply pressure or use any sharp objects when cleaning the pressure transducers as damage may
result to them. These are the two white plastic discs on the front that have contact with the Day Cassette.
Carry out all cleaning for other parts of the system according to specified procedures for each device.