Instructions for Use
Turn on the pump with On/Off switch (13). at the rear and On/Standby button (2). Press the Soft Key ‘START’.
The cassette loading selection menu appears.
The Cassette Loading menu can also be accessed by pressing either the Day Cassette or Patient Cassette button on the keyboard.
To use
Single only mode, only load the Day Cassette and connect the Single tubing, REF No. 1002052.
To use both Inflow and Outflow (
double pump mode), load the Day Cassette and the Patient Cassette. If a RF generator is to
be used in the procedure, a Double Pump RF Patient Cassette REF No. 1003059 must be used. This has a third suction port for
the RF Probe. If a RF generator is not to be used, either a Double Pump RF Patient Cassette REF No. 1003059 or a Double Pump
Patient Cassette REF No. 1002059 can be used. See Cassette Loading illustrations below.
Once the cassette has been correctly loaded and the corresponding LEDs are lit, `OK’ will be visible on the Display next to a Soft
Key. Press the Soft Key `OK’ to continue to the procedure selection menu. Pressing the Soft Key will return to the cassette
loading menu.
WARNING: Before and after opening, carefully examine the sterile packages for the cassettes and or Single Tubing. If any
damages are detected, the sterile barrier is broken and the product may not be used as sterility cannot be guaranteed.
Also, make sure that the expiry date not has been passed. If it has, the sterility and/or the function cannot be guaranteed.
The procedure selection menu provides four options for joint type: 1) Shoulder 2) Knee 3) Hip 4) Small Joint (i.e. wrist, elbow and
ankle joints). Each of these procedure types has specific user default settings for pressure, outflow, shaver suction and SmartVision
®
Outflow Tracking on/off. Changing the user default settings is described in ‘Changing user default Settings’ (page 21).
Select the appropriate joint type by moving the arrow with the Soft Keys or . When the arrow is adjacent to the appropriate
joint type, press the Soft Key ‘OK’. Pressing the Soft Key will take you back to the cassette loading menu.
The operator must be experienced in arthroscopic techniques. The user shall endeavor to remain current with advances in orthopedic
procedures. Additional training on the use of Double Pump RF from a Medical Vision AB representative is recommended.
Starting up
In Extended Setup, Quick Start can be enabled. Quick Start shortcuts most steps when setting up the Double Pump RF.
After loading the cassettes as described in the Load Cassette(s) section, a selection of the four procedure settings appear. After
selecting, just press Run/Stop to start liquid flow. When the pump is stopped it is possible to reach the cassette loading menu by
pressing “Day Cassette” button. Then, the Start and Setup menus can be reached by pressing Soft Key .
Quick Start
Load the cassette(s)
Procedure selection
Operator Training Requirements
See illustrations on pages 11 and 12.