spirodoc

 

 

 

SPIRODOC

 

 

 

cod. 980156 

 Rev 2.4   

 

EN 28/46 

 

User manual 

After the acoustic signal 

exspire slowly

 as much as air as possible and 

inspire slowly 

as much air as possible. 

To end the test touch the OK icon. 
Follow the indications on the display carefully in order to carry out the test properly. 
 

 

WARNING

 

To 

perform  a  test  in  the  service  menu  with  the  item  “PARAMETER  setting”,  at  least  one  parameter  related  to  this  test  must  be 

activated/chosen, otherwise the icon will be disabled. 

 

2.11.3 

MVV Test 

 

Start the test by carrying out a series of forced inspirations and expirations with the maximum possible amplitude. The suggested frequency is 30 
breaths per minute. The test will end automatically after 12 seconds. 

 

WARNING

 

To perform a test in the service menu 

with the item “Set parameter”, at least one parameter related to this test must be activated/chosen, 

otherwise the icon will be disabled. 

 

 

WARNING

  

The disposable mouthpiece and the disposable turbine must be replaced after a single patient test session. 

 

2.11.4 

POST test, after drug administration 

 

WARNING

  

To carry out a POST test it is necessary to have carried out at least one PRE FVC  test the same day; it is not possible to do a POST test 
on the PRE VC or MVV tests; it is however possible to do a POST VC or MVV test if the database already contains at least one PRE test 
carried out on the same day. 

To carry out a POST test please access to the spirometry area touching the 

 icon on the main screen and subsequently touching the 

 

icon. 
 
A  POST  test  is  a  spirometry  test  following  the  administration  of  a  drug  of  some  kind,  usually  a  bronchodilator.  The  sign  POST  is  shown  on  the 

screen  of the  device  (top  right)  during  the  following  tests.  On  the RHS  of  the  screen  there  is the  icon 

  which  enables  you  to  see  all of  the 

results from the PRE test as well as the relevant predicted values. The following tests made by the patient show the following parameters: 
 

 

Those values related to the test performed 

 

Those values related to the best PRE test performed by the same patient the same day.(that is in the same test session) 

 

The percentage variation between the PRE and POST values (in the CHG column)  

 
It is not possible to perform a POST test with a patient whose PRE testing was not carried out on the same day.  
If  during  a  POST  session  a  new  patient  is  inserted  or  another  is  recalled  from  the  archive  the  device  will  automatically  exit  the  current  POST 
session. 
 

2.12 

Viewing the spirometric results 

 
Following a FVC test, the spirometry test results are shown. The first screen displays: 
 

 

A Flow/Volume graph of the Forced Vital Capacity 

 

All the parameters selected (with the percentage change compared to the predicted values) in a configuration menu. 

 
By scrolling on the right hand side of the screen it is possible to view all the parameters next to the chosen predicted values. 
 

2.12.1 

Spirometry test interpretation 

Spirometry test interpretation is based on the Forced Vital Capacity (FVC) test. The test interpretation is indicated with one the following messages: 
 

 

Normal Spirometry 

 

Mild obstruction 

 

Moderate obstruction 

 

Moder severe obstruct. 

 

Severe obstruction 

 

Very severe obstruction 

 

Mild Restriction 

 

Moderate Restriction 

 

Moder Severe Restriction 

 

Severe Restriction 

 

Very severe Restriction 

 

Obstruct. + Restrict 

 
For a 

POST test the messages are the same but instead of dealing with an “obstruction” the POST test refers to a “restriction”. 

 
Through the use of a mathematical analysis applied to certain indices and parameters calculated in the FVC test , the 

SPIRODOC

 is capable of 

producing a list of quality control comments useful to assess the quality and reproducibility of the manouvers performed. 
 
The quality control check assigns a letter for the current spirometry session as described below: 

Содержание spirodoc

Страница 1: ...spirodoc SPIRODOC cod 980156 Rev 2 4 EN 1 46 User manual User manual rev 2 4 Issue date 10 04 2014 Approval date 10 04 2014...

Страница 2: ...23 2 7 Patient Data 24 2 7 1 Inserting data of a new patient 24 2 7 2 Patient data modification 24 2 8 Visualization of memory data 24 2 8 1 Database research modality 24 2 8 2 Visualization of databa...

Страница 3: ...mir spirometry com MIR accepts no responsibility for any loss or damage caused by the user of the device due to instructions contained in this Manual and or due to incorrect use of the product Please...

Страница 4: ...e a spirometry test is made The user is responsible to assess both the mental and the physical condition of the patient in order to perform a proper test furthermore in the evaluation of test results...

Страница 5: ...r the correct functioning and possibly cause damage The presence of any impurities such as hair sputum threads etc within the body of the turbine sensor may seriously compromise measurement accuracy 1...

Страница 6: ...with the correct operation of the device For this reason certain minimum clearances a few meters should be observed when high frequency appliances such as a TV radio portable phone etc and other elec...

Страница 7: ...rly aligned If the leads are inverted an inverse polarity connection may provoke acid leakage overheating smoke breakage and or fire Do not connect the battery pack leads to a wall socket or to the ca...

Страница 8: ...tandard this product and its component parts are of type BF and therefore protected against the dangers of direct and indirect contact with electricity 1 5 4 Warning symbol for the USB To connect to o...

Страница 9: ...ved receive an explanation of the ESD warning symbol and training in ESD precautionary procedures the electrostatic discharge is defined as an electric charge at rest It is the sudden flow of electric...

Страница 10: ...or alternatively to the personal best values from the clinical history of the subject The personal best values can vary considerably from the predicted values which are taken from healthy subjects SPI...

Страница 11: ...nt system Bi directional digital turbine Temperature sensor semiconductor 0 45 C Measurement principle Infrared interruption Volume range 10 L Flow range 16 L s Volume accuracy 3 or 50 mL Flow accurac...

Страница 12: ...to administer to the patient O2 percentage of oxygen administered to the patient before test L min SPO2 Base SPO2 base value for the SPO2 and ODI tests BPM Base BPM base value for the SPO2 and ODI te...

Страница 13: ...es SPO2 start 6MWT no SPO2 end 6MWT no BPM end 6MWT no SPO2 Base 6MWT no BPM start 6MWT no T Baseline 6MWT no T Walking 6MWT no T Recovery 6MWT no Distance 6MWT no T2 SPO2 T2 SPO2 6MWT yes T4 SPO2 T4...

Страница 14: ...ecial alarm frequency in the case of either SpO2 or cardiac pulse going outside of the programmed levels of alarm Beep with special alarm frequency during oximetry measurement in the case of a low bat...

Страница 15: ...llows access to all functions by simply touching the display The controls on the touch screen change dynamically based on the functions performed To access a specific function touch the corresponding...

Страница 16: ...ith the symbol This image indicates that the battery pack is fully charged 6 indicators A drop of the battery pack charge is displayed with a reduction of the indicators 2 4 Main screen On the main sc...

Страница 17: ...via Bluetooth connection To view the pletismographic curve in real time while performing an oximetry test To check the alarms and alarm thresholds during oximetry testing To check the alarms and alarm...

Страница 18: ...egin There are four phases as follows Touch three times the top left hand point of the display Touch three times the top right hand point Touch three times the bottom right hand point Touch three time...

Страница 19: ...a menu appears to set the day and time that the device will turn on and off If turn ON mode was previously enabled and the user turns on the device at a different time from that of the programmed tim...

Страница 20: ...beep during oximetry testing WARNING The heart rate tone beep is always disabled during sleep oximetry testing please view paragraph 3 13 2 Default alarms This function allows to restore all default...

Страница 21: ...alibration allows to insert new calculated values to perform a new calibration A password is required to access this option insert the following password by touching the numbers starting from left to...

Страница 22: ...he item Configuration and insert the following password 1 2 2 3 3 3 The service menu will display the following items Doctor Patient Turn ON mode Patient data Oximetry setup Set questions Set symptoms...

Страница 23: ...ot be accepted This means that the device is not capable of correcting such a large calibration error In this case Check the correct functioning of the SPIRODOC with a new turbine and or Clean the tur...

Страница 24: ...ian Caucasian 100 Mexican American Oriental 100 Afro American Hong Kong Chinese 100 Other Giapanese 89 polinesian 90 North Indian 90 South Indian 87 Pakistani 90 African descendant 87 Aboriginal 85 Wh...

Страница 25: ...t specific date Test session in Patient Mode refers to spirometry PRE tests and oximetry tests performed within a 20 minute time span A new session is activated upon turning on the device if the time...

Страница 26: ...s can be seen on the screen to the right If a patient has instead performed a test in the current session and previous tests are archived the following screenshot is displayed 2 10 PC On line mode con...

Страница 27: ...tory parameters FIVC FIV1 FIV1 PIF are not of interest The optional initial inspiration phase can also be performed before inserting the mouthpiece in the mouth After inspiring slowly and deeply the f...

Страница 28: ...results from the PRE test as well as the relevant predicted values The following tests made by the patient show the following parameters Those values related to the test performed Those values related...

Страница 29: ...s shown Blow out all air in lungs Between tests the SPIRODOC checks the repeatability of the following parameters PEF repeatable when the difference between the two largest PEF is 0 67 L s VC repeatab...

Страница 30: ...message will appear Low battery level Touch the ESC icon to exit the test otherwise after a seconds will start the test In the event that a test is interrupted due to a complete battery discharge the...

Страница 31: ...case the icon will be shown as to then return to the previous icon when the two minutes are over For information concerning the proper setup of this function please refer to paragraph 3 6 WARNING A t...

Страница 32: ...6 minutes are up the device will automatically move on to the recovery stage and the timer will be reset to zero again Recovery stage The duration of this stage is entirely up to the doctor and it is...

Страница 33: ...agraph 3 13 1 To perform this test touch the from the main screen and subsequently select the test with the icon After approximately 5 minutes from the beginning of the test the SPIRODOC will automati...

Страница 34: ...The materials used for manufacturing the sensor are NATURAL LATEX PROTEIN FREE and are subject to biocompatibility tests WARNING The adult single patient sensor is ideal for patients weighing more th...

Страница 35: ...ximetry test walk test sleep oximetry send data via Bluetooth WARNING Spirometry test results do not include any automatic interpretation compared with the predicted values as no patient anthropometri...

Страница 36: ...Vista and Seven To avoid making any errors at this stage please read the Advanced section of the winspiroPRO User Manual carefully 3 4 Internal software upgrade SPIRODOC internal software can be upgr...

Страница 37: ...ne is still malfunctioning please replace it with a new one WARNING Do not carry out any cleaning procedures when using disposable turbines a new disposable turbine must be used for every new patient...

Страница 38: ...spiroPRO software For further information consult the winspiroPRO manual available on line within the software itself Spirometry test results are unreliable The turbine may contain dirt or foreign mat...

Страница 39: ...irometer Oximeter Brandname MIR Medical International Research Device name spirodoc Class IIa Complies with the Essential Requirements of directive 93 42 EC concerning Medical Devices and its amendmen...

Страница 40: ...ation If the product is utilised differently from the use described in the User s Manual If any alteration adjustment modification or repair has been carried out by personnel not authorised by the man...

Страница 41: ...spirodoc SPIRODOC cod 980156 Rev 2 4 EN 41 46 User manual ANNEX 1OXIMETRY TEST REPORTS EXAMPLES...

Страница 42: ...spirodoc SPIRODOC cod 980156 Rev 2 4 EN 42 46 User manual...

Страница 43: ...spirodoc SPIRODOC cod 980156 Rev 2 4 EN 43 46 User manual...

Страница 44: ...spirodoc SPIRODOC cod 980156 Rev 2 4 EN 44 46 User manual ANNEX 2SPIROMETRY TEST REPORT EXAMPLE...

Страница 45: ...for input output lines Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode Not Applicable Mains power quality...

Страница 46: ...OC The SPIRODOC is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the SPIRODOC can help prevent electromagnetic interfe...

Отзывы: