Maintenance
Repairs
GA 5752 3125 GB 10
29
7
Defect
Source of malfunction
Troubleshooting
Oversuction of device
despite protective system
Tapping unit mounted at an
angle
Operate tapping unit in a vertical
position only
Overflow protection device
contaminated
Clean overflow protection device
No foam inhibitor used
Use standard commercial foam
inhibitor
No hydrophobic filter used
Use hydrophobic filter
Gas probe does not fit into
the terminal unit
Wrong terminal unit selected Check the designation of the
terminal unit
Fine regulator is leaking
Seal is missing or defective
Have fine regulator checked
There is a leak in the housing
Accessories are not tightened Check fit of accessory
Tab. 5:
Troubleshooting
7.4
Repairs
The following may require repairs by the manufacturer or an authorised service partner:
• Liquid has penetrated the device.
• Performance has significantly decreased.
• Inexplicable notifications appear.
• Abnormal noises occur.
• Functional faults cannot be rectified according to the measures in chapter Malfunctions and
If defects are detected, the product may not be used any longer.
Make a note of the defects and the REF number on the type plate and inform your ATMOS
representative.
Observe the information in chapter Sending in the device [
7.5
Service hotline
+49 7653 689-0
7.6
Sending in the device
Remove and properly dispose of consumables.
Clean and disinfect the product and accessories according to the operating instructions.
Place used accessories with the product.
Fill in form QD 434 ‘Delivery complaint / return shipment’ and the respective
decontamination
certificate
.
This form is enclosed with each delivery and can be found at www.atmosmed.com.
The device must be well padded and packed in suitable packaging.
Place form QD 434 ‘Delivery complaint / return shipment’ and the respective
decontamination
certificate
in an envelope.
Affix the envelope to the outside of the package.
Send the product to ATMOS or to your dealer.