Initial operation
Mounting accessories
GA 5752 3125 GB 10
19
3
3.3.2
Connection of the mechanical overflow protection
11
22
33
44
Fig. 5: Connection of the mechanical overflow
protection
Connection of the mechanical overflow
protection
Plug the overflow protection device (1)
directly onto the tube connector (2) on the
housing of the tapping unit and press
upwards until it stops.
Attach the connection tube (3) to the tube
connector (4) of the mechanical overflow
protection device and connect it to the
designated tube connector of the septic
fluid jar.
For disassembly, first remove the
connection tube from the tube connector of
the mechanical overflow protection device.
Then remove the complete overflow
protection device with lid from the tube
connector of the tapping unit. Hold the
tapping unit steady with one hand while
doing so.
3.3.3
Connection of the hydrophobic bacterial and viral filter
11
44
33
22
Fig. 6: Connection of the filter
Connection of the hydrophobic bacterial
and viral filter
The connection tubes (1) and (2) are
plugged onto the tube connectors of the
filter (3).
Check the flow direction of the filter. The
inlet side must face the patient, and the
outlet must face the tapping unit.
Then plug the connection tube (1) onto the
tube connector (4) on the housing of the
tapping unit.
Connect the connection tube (2) with the
designated tube connector of the septic
fluid jar.