Cleaning and disinfection
Cleaning
GA 5752 3125 GB 10
25
6
6.2
Cleaning
6.2.1
General
NOTE
Use only all-purpose cleaners which are slightly alkaline (soap solution) and
contain surfactants and phosphates as the active cleaning agents.
In the event of heavily contaminated surfaces, use concentrated all-purpose
detergent.
CAUTION!
Improper cleaning can cause property damage!
Residues of physiological saline solutions (e.g. sodium chloride) can attack the
surfaces of the product.
Remove residues of physiological saline solutions with a cloth slightly dampened in
clean water. Then dry the product with a dry, lint-free cloth.
CAUTION!
Improper cleaning can cause property damage!
Do not spray cleaning agent directly into the joints or gaps and never use a high-
pressure cleaning unit!
6.2.2
Cleaning procedure
Use the correct dose of all-purpose detergent with water for the degree of surface
contamination and in accordance with the instructions of the detergent manufacturer.
Thoroughly wipe off the product with a soft cloth slightly dampened in an all-purpose detergent
solution.
Ensure that the product is free of contamination and encapsulated particles of grime.
Thoroughly wipe off the product with a soft cloth slightly dampened in clean water.
Ensure that the product is free of detergent residues.
Dry the product with a dry, absorbent and lint-free cloth.
This will help to reduce pathogen growth on the product's surface.
Wipe disinfect the product after every cleaning process.
6.3
Disinfection
6.3.1
General
NOTE
In the event that product surfaces are very dirty, carry out an additional cleaning
procedure before disinfecting.
DANGER!
Reduced performance!
Only clean the product by wipe disinfection.
Ensure that no disinfectants enter the unit.
Check the functionality of the product after each disinfection.