Maintenance
General
28
GA 5752 3125 GB 10
7
7
Maintenance
7.1
General
Maintenance, repairs and periodic tests may only be carried out by persons who have the
appropriate technical knowledge and are familiar with the product. To carry out these measures,
the person must have the necessary test devices and original spare parts.
ATMOS recommends: Work should be carried out by an authorised ATMOS service partner. This
ensures that repairs and testing are carried out professionally, original spare parts are used and
warranty claims remain unaffected.
DANGER!
Health hazard!
The product is used in the treatment of patients. Any restriction of performance of
the product can result in complications during treatment.
Check proper function of the product each time it is reconditioned.
Users need not carry out any maintenance work beyond the usual daily cleaning and care.
7.2
Periodic tests
Observe the country specific requirements regarding periodic tests.
7.3
Malfunctions and troubleshooting
Defect
Source of malfunction
Troubleshooting
• No or low vacuum
• No or reduced flow rate
• Regulation of flow rate not
possible
Fine regulator not connected Attach the fine regulator in
accordance with the operating
instructions
The connection tube is not
connected to tapping unit
Connect connection tube
according to operating
instructions
Connection tube too long
Shorten the tube to a maximum
length of 50 cm
Connection tubes collapse
Use special connection tubes
(vacuum proof up to -95 kPa)
Full septic fluid jar; overflow
protection system closed
Empty/replace septic fluid jar;
replace overflow protection
system
Oversuction of hydrophobic
filter
Replace hydrophobic filter
Seal damaged
Replace seal
Suction system is leaking
Check suction system
Central gas supply system
failure
Contact Technical Service
Probe connection loose
Fine regulator defective
Adjusting screw defective