U5002-1, 7.0
37
XIII.
System Compliance Statements (continued)
IEC 60601-1-2:2004(Ed 2.1) Table 206 Requirements:
Recommended separation distances between portable and mobile RF communications equipment and the HDMD.
The Equipment is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or
the user of the Equipment can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the Equipment as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output power of
transmitter W
Separation distance according to frequency of transmitter M
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.24
0.1
0.37
0.37
0.74
1
1.17
1.17
2.34
10
3.69
3.69
7.38
3,5
V
1
3,5
E
1
80 MHz to 800 MHz
7
E
1
800 MHz to 2,5 GHz
3,5
V
1
3,5
E
1
80 MHz to 800 MHz
7
E
1
800 MHz to 2,5 GHz
3,5
V
1
3,5
E
1
80 MHz to 800 MHz
7
E
1
800 MHz to 2,5 GHz
Содержание HDMD