1.7 Product parts and accessories
The Veterinary Temp-pulse Oximeter is composed of host, USB cable, tongue-clip
SpO2 probe and reusable body cavity temperature probe for animal. The host is
composed of display screen, control buttons and so on. The accessories include
SpO2 probe, reusable body cavity temperature probe for animal and USB cable.
1.7.2
Accessories
The measurement of pulse oximetry is achieved by calculating the change in the
absorption rate of transmitted light during capillary pulsation of animal limbs, tongue,
and earlobe, etc., which is a continuous, non-invasive measurement method of
hemoglobin oxygen saturation ( Functional saturation). Since oxyhemoglobin and
deoxyhemoglobin show different absorption rates for two specially selected incident
lights with different wavelengths, different oxyhemoglobin saturation can be
determined, ie, FO2Hb.
The company's Veterinary Temp-pulse Oximeter uses two transmitted light sources
with different wavelengths, namely about 660 nm for red light and about 905 nm for
infrared light. One side of the sensor contains two LEDs, and the other side contains
a photoelectric detector. The two luminous tubes emit lights alternately. The light is
weak when the pulsation is measured by the photoelectric sensor, and the light is
stronger between two pulsations, thus the reduced value is the light intensity
absorbed by the pulsatile arterial blood. In this way, the light absorption ratio of the
two wavelengths can be calculated, thereby the oxygen saturation of the arterial blood
can be accurately measured within the range of 70%~100%, obtaining the "pleth"
waveform and the pulse rate value.
1.8 Principle of Measurement
Fractional saturation FO2Hb= X100
O2Hb
RHb
+
O2Hb
+
MetHb
+
COHb
Functional oxygen saturation
=
X100
O2Hb
RHb
+
O2Hb
1) As the measurement is taken based on the pulsation of the arteriole, the tester's
blood should ensure a steady flow. For an animal subject with weak perfusion due to
shock, low ambient/body temperature, massive haemorrhage, or use of vascular-
contracting drug, the SpO2 waveform will decrease. In this case, the measurement
will be more sensitive to interference.
2) For those drugs with a substantial amount of staining dilution such as blue, indigo,
green, and acidic indigo methylene, carbon monoxide hemoglobin, methionine,
thiosalicylic hemoglobin and some icterus problems, the usage of the device for
measurement of SpO2 may be inaccurate.
3) Drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also be a
major factor blamed for serious error of SpO2 measurement.
4) The SpO2 value serves only as a reference value for judgement of anemic anoxia
and toxic anoxia in related animals. It can also be used as a measurement of pulse
oximetry for severe anaemic animals.
1.9 Clinical Restrictions
1.7.1
Host
Figure 2 Structure of host
Control
button
Display
Screen
Cable
Probe
Connector
Connector
Probe
Cable
Device connector
cable
Pulse oximetry
probe & Temperature
probe connector
Figure5
USB Cable
Figure3
Tongue-clip SpO2 probe for animal
Figure4
reusable body cavity temperature
probe for animal