Phenomena
Possible Causes
Solutions
Abnormal on-off
of Veterinary Temp-
pulse Oximeter
(display screen
and LCD emitting
tube don’t lighten
normally)
The power button isn’t
pressed in place
Battery is exhausted
Re-press the power button in place, and
keep 1-2 seconds
Charge the battery in time.
No display on screen,
but the emitting
tube
of LED light
lightens
normally.
Display screen damage or
damage of the connection
point to the display screen
SpO2 and PR
value are
shown as "--"
The sensor position
is incorrect
Readjust the probe position
Program initialization
After connecting the animal subject, if the
waveform of the pulse wave shows, it indicates
the program is carrying out the initialization
calculation. Please keep the animal still and
stable, after which the values can be displayed
after the initialization calculation is completed.
Poor perfusion problem
If the pulse amplitude has no display on screen
Please
Adjust the measurement position;
Warm
the measurement position
.
,
Poor perfusion problem
The probe is malfunctional and
the LED tube is not illuminated
or the receiving
tube is
damaged.
Probe extension cable (if
any) is malfunctional.
Replace the original probe extension cable
or contact an authorized distributor
Flashing of
SpO2 value
1)The SpO2 value is greater
than or equal to the prompted
upper limit value
The SpO2 value is less than
or equal to the prompted lower
limit value
2)
1) Pay attention to whether the physiological
state of animal subject is normal. If there is
any abnormality, please consult a professional
medical staff.
2) Through the menu settings, you can modify
the upper and lower limits of the prompt, as
shown in Table 1.
If the "Alm" function is turned on, the SpO2
value flashes with a sound of “Beep-Beep”.
The sound can be turned off by turning off the
"Alm" function, as shown in Table 1.
17
18
Section 3- Troubleshooting
This section explains how to troubleshoot the Oximeter. The table below lists some
possible abnormal conditions and possible faults. We recommend the corresponding
action according to these possible faults. See Table 3 below for details.
Table 3
:
Troubleshooting Guide
The
temperature
reading is
displayed
as "--.-"
Reusable body cavity
temperature probe for
animal is not inserted.
Insert reusable body cavity temperature probe for animal
to re-measure.
Reusable body cavity
temperature probe for
animal malfunctions.
The device display specification is exceeded. If the
“Alm” function is turned on, there is also a “Beep-
Beep--” sound, which can be turned off by turning off
the “Alm” function.
Temperature value goes
beyond the display range
of the device
The temperature measurement function is malfunctional.
If the “Alm” function is turned on, there is also a “Beep-
Beep--Beep-” sound,
which can be turned off by turning off the “Alm” function.
Please contact an authorized distributor.
Failure of temperatrure
self-test
Replace the reusable body cavity temperature probe for
animal or contact an authorized distributor.
Pulse rate
value
flashing
1) Pulse rate value is greater
than or equal to Upper limit
of the prompt;
2) The pulse rate value is
less than or equal to The
lower limit of the prompt.
1) Pay attention to whether the physiological state of the
tested animals is normal. If there is any abnormality,
please consult a professional medical staff.
2) Through the menu setting, you can modify the upper
and lower limits of the prompt. See Table 1 for details.
If the “Alm” function is enabled, the pulse rate value
flashes with “
” sound, which can be turned
off by “Alm” function. Turn off the sound prompt, see
Table 1 for details.
Beep-Beep--
Flashing of
temperature
value
The temperature value
is higher than or equal to
the prompted upper limit
value
The temperature value is
lower than or equal to the
prompted lower limit value
1) Pay attention to whether the physiological state of
animal subject is normal. If there is any abnormality,
please consult a professional medical staff.
2) Through the menu settings, you can modify the
upper and lower limits of the prompt, as shown in
Table 1.
3) If the "Alm" function is turned on, the temperature
value flashes with a sound of “Beep-Beep”. The sound
can be turned off by turning off the "Alm" function, as
shown in Table 1.
Failure of
automatic
shutdown
External Temp-probe is still
working
Pull out the external temp-probe
Inaccurate
Temp.
Measurement
Damage in receiving tube
or other parts.
Contact authorized distributors
The skin surface sensor is
not firmly stuck on the skin
surface.
Please use a medical tape to tape the sensor on the
measurement position.
Wrong position of Temp-
Probe
Place the Temp-Probe on proper position according to
the instructions
for
use
No enough measuring time
Maintain the correct measurement method and read
after 90s.
Contact authorized distributors
Contact authorized distributors
Phenomena
Possible Causes
Solutions
Temperature value goes
beyond the measurement
range.
1) When the temperature value exceeds the measurem-
ent specifications, the user shall be prompted.
2) If the "Alm" function is turned on, the temperature
value flashes with a sound of “Beep-Beep”. The sound
can be turned off by turning off the "Alm" function, as
shown in Table 1.