Section 1- Overview

1.1

 

Intended Use

The  Veterinary  Temp-pulse  Oximeter  is  intended  for  non-invasive  detection  and 
measurement  of  arterial  oxygen  saturation(SpO2),  pulse  rate respiration

 

rate  and 

temperature of animal. 

1.2

 

About the Veterinary Temp-pulse Oximeter

The  device  contains  a  dual  light  source  (red  LED  and  infrared  LED)  and  a  photo 
detector. Bone, tissue, pigmentation and venous vessels normally absorb a constant 
amount of light over time. The arteriolar wall normally pulsates and absorbs variable 
amounts  of  light  during  the  pulsations.  Functional  oxygen  saturation  (SpO2)  can  be 
calculated from the ratio of absorbed light. Since the measurement of SpO2 depends 
on light source from the device, a strong light source in the surrounding environment 
will affect the measurement accuracy.
If  the  animal  temperature  is  measured,  an  external  temperature  probe  specified  by 
Medlinket must be connected to the host. Refer to Table 2 for details.

1.3

 

Identification of Panel and Symbols 

A complete description of all buttons, symbols, controls, displays and indicators.

Figure 1: Front Panel

 

and

 

L

eft

 

P

anel

 

o

f

 

Veterinary

 

T

emp

-

p

ulse

 

O

ximeter 

1— Menu button/Power button

              

6— Waveform Display

2— %SpO2 Display

                                   

7— Bar graph (The Pulse Amplitude Indicator)

3— Battery indicator

                                  

8— Display direction switch 

4— Temperature Display

                          

9— Accessory connector

5— Pulse Rate Display (bpm)

1.4 Technical Specifications

Battery

Temperature

Note: This temperature measurement function can only be achieved with a connection to our 
company’s

 

temperature probe. See Table 2 in this instructions

 

for

 

use for details.

Operating Current

Pulse and oxygen saturation

SpO2 Measurement Accuracy

SpO2 Resolution

Reminder

90% to 100% range: ±2%;
70% to 89% range: ±3%
<70%: unspecified 

Low-battery indicator. When the battery is low, 

the device will be turned off after the voice of 

“beep-beep” 

1%

Method

Pulse Rate Resolution

Optical output power

Pulse Rate

 

Measurement Accuracy

Pulse Rate Measurement Range

LED Wavelength

Updating time 

Average data time 

Dual wavelength LED transmitting

 

tube

1

 

bpm

Red: approximately 660nm; 

Infrared: approximately 905nm

1 sec

30 to 300bpm

±

3

 

bpm

Less than 15mW

8 sec 

SpO2 Measurement Range

35% to 100%

Less than 55mA

Only Temp function works

Less than 40mA

SpO2 and Temp function 

work together

Less than 60mA

Only SpO2 function works

Measurement Range & Accuracy

Temperature Resolution 

77°to 113°F (25

 to 45

℃): ±

0.2

±

0.1

Battery specification

3.7V Rechargeable lithium battery

2

4

1

2 3

5

8

6

7

9

Respiration

 

Rate

(

RR

Measurement

 

Range

 

Respiration

 

Resolution

 

Respiration

 

Measurement

 

Accuracy

6rpm

-

70rpm

1

 

rpm

±

2

 

rpm

Содержание AM-806VB

Страница 1: ...Fax 86 755 61120055 Email kefu medxing com Website www medxing com Shenzhen Med link Electronics Tech Co Ltd AM 806VB REV A 1 Revision date October 24 2019 Veterinary Temp pulse Oximeter Instructions...

Страница 2: ...t sites that are not well perfused To prevent this situation please check application sites of animal subject frequently If the initial skin temperature does not exceed 35 C the operation of the senso...

Страница 3: ...ing inspection 07 1 12 Packaging 07 Chapter II Operation 08 2 1 Installation and confirmation 08 2 2 Operation and use 10 2 3 Security 11 2 4 Function settings 12 2 5 Menu Settings 13 2 6 List of appl...

Страница 4: ...2 Display 7 Bar graph The Pulse Amplitude Indicator 3 Battery indicator 8 Display direction switch 4 Temperature Display 9 Accessory connector 5 Pulse Rate Display bpm 1 4 Technical Specifications Bat...

Страница 5: ...rement please use the reusable body cavity temperature probe for animal to measure the same position for at least 90 seconds Medical high frequency signals or defibrillator interference can cause erro...

Страница 6: ...the two wavelengths can be calculated thereby the oxygen saturation of the arterial blood can be accurately measured within the range of 70 100 obtaining the pleth waveform and the pulse rate value 1...

Страница 7: ...ease contact the Technical Service Department You can contact by Phone 86 755 61120085 Fax 86 755 61120055 Email user07 med linket com 1 11 Unpacking and Inspection 1 12 Included in the package Sectio...

Страница 8: ...ning properly or not Caution 2 2 General Operation Preparation 1 Please make sure that the host battery is fully charged If the battery is low please plug in the USB cable to charge 2 Press the power...

Страница 9: ...pressure cuff or intravenous injection Make sure the light source is from the internal transparent silica gel Any other light source may cause incorrect SpO2 measurement and venous pulsation Strong l...

Страница 10: ...completing the setting the interface will indicate OK Move the cursor to select Exit by short pressing the button and then long press the button to return to the testing interface Setting of SpO2 upp...

Страница 11: ...em of Exit to return to the testing interface Return to the testing interface Warning Uncomfortable or painful feeling may appear if using the device ceaselessly especially for animals with blood micr...

Страница 12: ...sound which can be turned off by turning off the Alm function Temperature value goes beyond the display range of the device The temperature measurement function is malfunctional If the Alm function is...

Страница 13: ...ng time please charge it at least once every 3 months to maintain the battery The best storage environment of the device is between room temperature 10 C and 40 C with humidity not higher than 80 Plea...

Страница 14: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 O indicates that the content of the toxic and hazardous substances in all homogeneous materials of the part is subject to the...

Страница 15: ......

Страница 16: ...stomer name Tel Address Product Name Product REFNO Purchase Date Cityof Purchase FailureDescription Product maintenance card This joint is reserved for customers with this card repair and return Custo...

Отзывы: