The Veterinary Temp-pulse Oximeter manual is intended to provide information for
proper operation and maintenance. General knowledge of monitoring and
understanding of the features and functions of the Veterinary Temp-pulse Oximeter
are prerequisites for proper use. Please read the instructions for use carefully before
using this device.
Please operate the device in strict accordance with the operation procedures
described in the instructions for use. Failure to follow these instructions may cause
improper measurement, equipment damage, injury of animal subject and so on. The
manufacturer is NOT responsible for the insecurity, performance instability, injury of
animals, device damage, etc. due to user’s negligence of the instructions for use.
The Veterinary Temp-pulse Oximeter is a medical device, and can be used
repeatedly.
Warning
Foreword
!
Explosion hazard. Do not use the Veterinary Temp-pulse Oximeter in the
presence of flammable anesthetics or gases.
The device is limited to be used by one animal at the same time.
Do not spray, pour, or spill any liquid on the Veterinary Temp-pulse Oximeter
together with its accessories, connectors and switches.
Repeated use of the sensor must be applied on a new position every 4 hours
during the test. Because animal subject’s skin condition affects the ability of the
skin to tolerate sensor placement, it may be necessary to change the sensor site
more frequently concerning some animals. If skin integrity changes due to the
placement of the sensor, move the sensor to another site.
With the increasing of the ambient temperature, skin of animal subject may be
severely burned after prolonged sensor application at sites that are not well
perfused. To prevent this situation, please check application sites of animal
subject frequently. If the initial skin temperature does not exceed 35 ° C, the
operation of the sensor will not cause the skin temperature to exceed 41 ° C.
Inaccurate
respiration
rate
(
RR
)
measurements
may
be
caused
by
improper
placement
or
alignment
,
low
arterial
perfusion
,
excessive
motion
and
cardiac
arrhythmias.
After the sensor is removed from the animal subject, data and waveform may still
be displayed on the oximeter due to external light interference. This does not
mean that the probe malfunctions, but the value displayed at this time cannot be
used as a basis for clinical diagnosis.
Portable electrical equipment and mobile RF communication equipment may
affect the usage of the product.
Veterinary Temp-pulse Oximeter must not be disposed as unsorted normal waste
but be classified and managed. Please contact an authorized representative of
the manufacturer for correct information concerning the disposal of the product.
Please refer to the related literature about the clinical restrictions and
precautions.
This device is not intended for treatment.
The LCD panel contains toxic chemicals. Please do not put the damaged LCD
panel in the mouth or swallow it.
Do not modify this device without authorization of the manufacturer.
Latex Content Statement
All parts of the Veterinary Temp-pulse Oximeter and accessories that are in contact
with the animal subject are free of latex.
About This Manual
This manual is provided for the operator or the user to read carefully after which
they can use the oximeter correctly.
Product model: AM-806VB
Product name: Veterinary
Temp
-
p
ulse
O
ximeter
Trade name: Pulse
Oximeter
Registration / production address: 4th Floor, Building A, Ying Tailong Ind Park,
Dalang South Road, Longhua District, Shenzhen
Product information