2.0 Product Accessories and Options
WARNING: POTENTIAL ELECTROSTATIC CHARGING HAZARD
In hazardous locations:
●
Avoid building up static charge on case of data protector and plastic accessories.
●
Use damp rag to wipe data protector and plastic accessories to avoid static build up.
Nose Caps and Pressure Connections
There are several different pressure connection options for the MEAS LT submersible level and
pressure transducer product line.
The LTA features a ported nose cap (Delrin) with #8-32UNC-2B threaded-hole where weight
accessories can be attached.
The LTB features a wide mouth non-fouling cage protecting a 2.75” PTFE flexible diaphragm.
The LT1 features a ¼” – 18 female NPT pressure connection.
The LT5 features a ¼” – 18 male NPT pressure connection.
The LT7 features a ½” – 18 male NPT pressure connection.
WARNING:
Caution must be exercised when inserting a screw into the nose cap as the
maximum insertion length should not exceed 0.175”.
810 SuperDry™ Vent Filter or 815 Aneroid Bellows
All submersible level transducers and pressure transducers with molded cable attachment require an
optional protective barrier that guards against moisture buildup in the cable vent tube. The 810
SuperDry™ vent filter and the 815 Aneroid Bellows are guaranteed to operate maintenance free for
one year. The 815 Aneroid Bellows can be used on submersibles with accuracies of ±0.25% FSO.
These barriers ensure reliable operation and long life as they protect sensitive electronic components
from mildew and prevent the formation of a liquid column in the vent tube. Any such liquid column
directly affects the instantaneous calibration of the transducer. Transducers with accuracy better than
+1 – 0.25% should only be protected with the 810 vent filter in order to maintain this accuracy.
Sacrificial Anodes
Our sacrificial anodes are made from a special zinc alloy formulated to guarantee continued
effectiveness over long periods. Because the anodes are 95% galvanic, they will not corrode unless
there is an electrolytic demand. The anode maintains a high driving potential throughout its 12 month
life, is self-sloughing and always exposes new zinc for the best possible protection. For those
applications where cable buoyancy is a problem, the sacrificial anode can be substituted for hanging
weights.
Sacrificial Anodes provide cathodic protection against galvanic corrosion for our submersible pressure