Cleaning:
Step 1:
Holding the cable 6 inches (150 mm) from the transducer, immerse the unit in the bowl
containing the mild detergent and stir for 20-30 seconds. Remove and rinse in the bowl
containing the fresh water, using the same stirring motion used in the mild detergent.
Rinse and wipe dry.
Step 2:
Holding the body of the transducer with one hand so that you are looking at the retaining
screen protecting the sensor, carefully remove the sensor nose piece by simply
unscrewing it from the transducer body.
Do not touch the sensor diaphragm with your
finger or any other object. Also, do not try to dry the inside portion of the transducer, as
you risk damaging the pressure sensor
.
Step 3:
Place the transducer in a vertical position with the sensing end facing downward in the
bowl containing a mild cleaning solution for approximately 15-20 seconds. Rinse in the
bowl containing clean water and wipe dry the external casing only. Place the protective
screen in the same solution for 15- 20 seconds, rinse and wipe dry.
Step 4:
Holding the transducer in a vertical position so that you can see the face of the sensor,
screw the protective nose piece back into place.
Cleaning the Lightning Protection Data Line Protector
WARNING
: POTENTIAL ELECTROSTATIC CHARGING HAZARD
In hazardous locations:
●
Avoid building up static charge on case of external surge protector and plastic
accessories.
●
Use damp rag to wipe data protector and plastic accessories to avoid static build up.