ANI Monitor V1 - Continuous analgesia monitoring system
39
MD/PRD/IN16.ANIV1 V.14 - 12 MAR 2020
Caution:
do not autoclave the acquisition device or monitor. Autoclaving will seriously
damage both components.
Avoid liquid ingress with the connection cables. This could interfere with acquisition
performance.
11.2
Maintenance
The ANI Monitor V1 is designed so that no periodic maintenance is required. However,
leakage current should be systematically checked, after spillage of blood or solutions, or after
a major surge in the house electrical system or at least once a year.
Preventive maintenance is recommended once a year: cable check, integrity of labels, system
check and leakage courant check according to the 62353 standard (for more details, refer to
the technical manual available from Mdoloris). However leakage current must be checked
systematically after every blood or liquid spill, or immediately after a major surge in the
electrical system.
Only Mdoloris Medical Systems qualified technicians and engineers or trained staff are
authorized to perform repairs and/or maintenance operations.
Only staff trained by Mdoloris Medical Systems can safely perform maintenance operations
(such as software updating or system recovery). However, the following elements may be
replaced or substituted by personnel untrained in technical maintenance (following the
manufacturer's instructions):
-
End User Cable;
-
Power cord;
-
Pole clamp.
12
Specifications, warranty and software license contract
12.1 Specifications
Electricity supply: 100/240 Volts
Consumption: < 32 W
Current: < 3 A