ANI Monitor V1 - Continuous analgesia monitoring system
14
MD/PRD/IN16.ANIV1 V.14 - 12 MAR 2020
Connect the black cable of the acquisition device to the monitor at the plug in the bottom left
hand corner of the monitor's front (figures 10 and 11).
3.4
Battery
The ANI Monitor V1 contains an internal battery. The battery means the system can keep
calculating the ANI while the patient is being moved or in case of main power supply failure.
The internal battery is not designed for long-term monitoring. The battery gauge has ten
graduations; each symbolizes around 10 min of battery life. As soon as the power supply is
interrupted, a message appears on the screen with a warning sound to ask the user to plug the
monitor back on the main power supply. The ANI index is still calculated during this time.
NOTE:
plug the ANI Monitor V1 on the main power supply before trying to switch it on.
ANI Sensor
Sensor cable
ANI Sensor V1
ANI-SENS-V1
ANI-SC-V1
ANI Sensor V2
ANI-SENS-V2
ANI-SC-V1
ANI Sensor V1 PLUS
ANI-SENS-V1PLUS
ANI-SC-V1PLUS
Figure 10
Figure 11
ANI-SC-V1
ANI-SC-V1PLUS