ANI Monitor V1 - Continuous analgesia monitoring system
17
MD/PRD/IN16.ANIV1 V.14 - 12 MAR 2020
Figure 15
Check for good ECG signal quality in the lower part of the screen. If there is no signal, check
the connection between the monitor and the electrodes. Try at least once to unplug and plug in
the connection again to recover the ECG signal. If a signal is displayed but does not look
right, check that the electrodes are properly placed and connected to the cable.
WARNING
:
always check in the upper left window that the signal quality is satisfactory
(green indicator). The ANI index will not be reliable if the signal quality is not good enough
(red indicator).
If the user thinks the ECG amplitude is too low, select
[Reset ECG]
to automatically
recalibrate the ECG signal acquisition.
5
Using the ANI Monitor V1 and setting parameters
5.1
ECG Capture
The lower window shows the ECG acquired by the monitor (figure 16). This ECG is filtered
of all technical and physiological artifacts (e.g. premature ventricular contractions).