Maxtec Handi+ R218P12 Скачать руководство пользователя страница 9

      866.4.Maxtec    www.maxtec.com

3

2.2 Calibrating the 

Handi+

 Oxygen Analyzer

Calibration of the instrument is necessary if the temperature of the gas stream 
changes by more than 3 degrees Celsius.
Changes in elevation result in calibration error of approximately 1% of reading 
per 250 feet. In general, calibration of the instrument should be performed 
when the geographic elevation at which the product is being used changes by 
more than 500 feet.  
In addition, calibration is recommended if the user is unclear when the last 
calibration procedure was performed or if the measurement value displayed is 
in question.
Calibrate the Handi+ to an oxygen source, which has a traceable certificate 
and/or USP certification, of either 20.9% oxygen, as found in room air, or 
100% oxygen. It is recommended that the oxygen source be documented at a 
pressure and flow similar to your clinical application. 

Note: 

Before beginning calibration the Handi+ must be in thermal equilibrium.  

You may also need to be aware of other factors, which affect device calibration 
values. For more information, refer to "Factors Influencing Calibration and 
Performance" on page 5 of this manual.

Note:

 

We recommend use of medical grade USP or >99% purity oxygen when 

calibrating the Handi+.

2.3 In Line Calibration (Configuration A)

1. Put the Handi+ in an upright position such that you can read the product label.
2. Connect a sample supply hose to a standard "T" piece.  The Maxtec "T" is 

precision-tapered to insure a tight connection with O-rings of the MAX-250 
oxygen sensor diverter.

3. Insert the Handi+ in the center position of the "T" piece.
4. Attach an open-ended reservoir to the end of the "T" piece. Then start the 

calibration flow of oxygen at 1-10 liters per minute.

Note:

 Six to 10 inches of corrugated tubing works well as a reservoir.

Note:

 A calibration oxygen flow to the Handi+ of 2 liters per minute is 

recommended to minimize the possibility of obtaining a "false" calibration value.
5. Allow the oxygen to saturate the sensor. Although a stable value is usually 

observed within 30 seconds, allow at least 2 minutes to ensure that the 
sensor is completely saturated with the calibration gas.

6. If the Handi+ is not already turned on, do so now by pressing the analyzer 

ON button

7. Press the “CAL”

button on the Handi+. The calibration gas value on the 

analyzer display should read 20.9 or 100 depending on the gas stream used.

Follow instructions for use

On/off Button

Calibration Button

Authorized Representative 
in the European Community

Lot code/Batch code

Drip Proof

Serial Number

Catalog Number

Содержание Handi+ R218P12

Страница 1: ...H Handi O P E R A T I N G M A N U A L I N S T R U C T I O N S F O R U S E R 2 1 8 P 1 2 R218M12 Rev J CAL...

Страница 2: ......

Страница 3: ...s from the date of shipment from Maxtec under normal operating conditions and provided that the Handi is properly operated and maintained in accordance with Maxtec s operating instructions Should Hand...

Страница 4: ...Improper use of this device can cause inaccurate oxygen readings which can lead to improper treatment hypoxia or hyperoxia Follow the procedures outlined in this user manual CAUTION Indicates a potent...

Страница 5: ...ation A 3 2 4 Direct Flow Calibration Configuration B 4 3 0 OPERATING THE HANDI 4 4 0 FACTORS INFLUENCING ACCURATE READINGS 5 4 1 Elevation Changes 5 4 2 Temperature Effects 5 4 3 Pressure Effect 5 4...

Страница 6: ......

Страница 7: ...blank The analyzer turns off after 1 3 minutes if no buttons are pressed Calibration Button Press the calibration CAL button to adjust the calibration value to reflect the known oxygen concentration T...

Страница 8: ...his oxygen analyzer ideal for portable oxygen analysis by qualified heath care professionals 2 0 CALIBRATING THE Handi 2 1 Before You Begin A protective film covering the threaded sensor face must be...

Страница 9: ...encing Calibration and Performance on page 5 of this manual Note We recommend use of medical grade USP or 99 purity oxygen when calibrating the Handi 2 3 In Line Calibration Configuration A 1 Put the...

Страница 10: ...ess the CAL button on the Handi The calibration gas value on the analyzer display should read 20 9 or 100 depending on the gas stream used 3 0 OPERATING THE HANDI To Check the Oxygen Concentration of...

Страница 11: ...e Handi will hold calibration and read correctly within 3 when in thermal equilibrium within the operating temperature range The device must be thermally stable when calibrated and allowed to thermall...

Страница 12: ...an error code occurs E03 No valid calibration data available Make sure unit has reached thermal equilibrium Press and hold the Calibration Button for three seconds to manually force a new calibration...

Страница 13: ...point on display located after the first digit Accuracy and Linearity 1 of full scale at constant temperature R H and pressure when calibrated at full scale Total Accuracy 3 actual oxygen level over...

Страница 14: ...on Dioxide 10 2 Helium 70 2 8 0 HANDI SPARE PARTS AND ACCESSORIES 8 1 Standard Replacement Parts and Accessories Part Number Item R218M12 Operation Manual Handi R218M12 001 Operation Manual Handi Inte...

Страница 15: ...calibration value Make sure that the calibration gas is connected to the Handi at 2 liters per minute for a minimum of 2 minutes prior to proceeding with calibration This 2 minute equilibration time...

Отзывы: