Maxtec Handi+ R218P12 Скачать руководство пользователя страница 11

      866.4.Maxtec    www.maxtec.com

5

NOTE:

 

If the Handi+ is used to measure the oxygen concentration with 

equipment using a heated or humidified gas stream, it is recommended that 
the Handi+ be placed upstream of the heater and/or humidifier. For more 
information, refer to "Factors Influencing Calibration and Performance" on 
page 5 of this manual.

For hospital and home care a new calibration is required when:

»  The measured O

2

 percentage in 100% O

2

 is below 97.0% O

2

»  The measured O

2

 percentage in 100% O

2

 is above 103.0% O

2

.

For ID testing (or optimum accuracy), a new calibration is required when:

»  The measured O

2

 percentage in 100% O

2

 is below 99.0% O

2

.

»  The measured O

2

 percentage in 100% O

2

 is above 101.0% O

2

.

 4.0 FACTORS INFLUENCING ACCURATE READINGS

4.1 Elevation Changes

»  Changes in elevation result in a reading error of approximately 1% of reading per 250 feet.   
»  A change in altitude greater than 500 ft will require sensor recalibration.
»  This device does not automatically compensate for changes in barometric pressure 

or altitude. If the device is moved to a location of a different altitude, it must be 
recalibrated before use (see section 2.2).

4.2 Temperature Effects

The Handi+ will hold calibration and read correctly within ±3% when in 
thermal equilibrium within the operating temperature range.  The device must 
be thermally stable when calibrated and allowed to thermally stabilize after 
experiencing temperature changes before readings are accurate. For these 
reasons, the following is recommended:

»  When used in a breathing circuit, place the sensor upstream of the heater.
»  For best results, perform the calibration procedure at a temperature close to the 

temperature where analysis will occur.

»  Allow adequate time for the sensor to equilibrate to a new ambient temperature.

4.3 Pressure Effect

Readings from the Handi+ are proportional to the partial pressure of oxygen. 
The partial pressure is equal to the concentration times the absolute pressure. 
Thus the readings are proportional to the concentration if the pressure is 
held constant. Flow rate of sample gas can affect pressure at the sensor in 
that back pressure at the sensing point may change. For these reasons, the 
following is recommended:

» Calibrate the Handi+ at the same pressure as the sample gas.
»  If sample gases flow through tubing, use the same apparatus and flow rates when 

calibrating as when measuring.

»   The  Handi+ oxygen sensor has been validated at pressures up to 2 atmospheres 

absolute. Calibration or operation above this pressure is beyond the intended use.

Содержание Handi+ R218P12

Страница 1: ...H Handi O P E R A T I N G M A N U A L I N S T R U C T I O N S F O R U S E R 2 1 8 P 1 2 R218M12 Rev J CAL...

Страница 2: ......

Страница 3: ...s from the date of shipment from Maxtec under normal operating conditions and provided that the Handi is properly operated and maintained in accordance with Maxtec s operating instructions Should Hand...

Страница 4: ...Improper use of this device can cause inaccurate oxygen readings which can lead to improper treatment hypoxia or hyperoxia Follow the procedures outlined in this user manual CAUTION Indicates a potent...

Страница 5: ...ation A 3 2 4 Direct Flow Calibration Configuration B 4 3 0 OPERATING THE HANDI 4 4 0 FACTORS INFLUENCING ACCURATE READINGS 5 4 1 Elevation Changes 5 4 2 Temperature Effects 5 4 3 Pressure Effect 5 4...

Страница 6: ......

Страница 7: ...blank The analyzer turns off after 1 3 minutes if no buttons are pressed Calibration Button Press the calibration CAL button to adjust the calibration value to reflect the known oxygen concentration T...

Страница 8: ...his oxygen analyzer ideal for portable oxygen analysis by qualified heath care professionals 2 0 CALIBRATING THE Handi 2 1 Before You Begin A protective film covering the threaded sensor face must be...

Страница 9: ...encing Calibration and Performance on page 5 of this manual Note We recommend use of medical grade USP or 99 purity oxygen when calibrating the Handi 2 3 In Line Calibration Configuration A 1 Put the...

Страница 10: ...ess the CAL button on the Handi The calibration gas value on the analyzer display should read 20 9 or 100 depending on the gas stream used 3 0 OPERATING THE HANDI To Check the Oxygen Concentration of...

Страница 11: ...e Handi will hold calibration and read correctly within 3 when in thermal equilibrium within the operating temperature range The device must be thermally stable when calibrated and allowed to thermall...

Страница 12: ...an error code occurs E03 No valid calibration data available Make sure unit has reached thermal equilibrium Press and hold the Calibration Button for three seconds to manually force a new calibration...

Страница 13: ...point on display located after the first digit Accuracy and Linearity 1 of full scale at constant temperature R H and pressure when calibrated at full scale Total Accuracy 3 actual oxygen level over...

Страница 14: ...on Dioxide 10 2 Helium 70 2 8 0 HANDI SPARE PARTS AND ACCESSORIES 8 1 Standard Replacement Parts and Accessories Part Number Item R218M12 Operation Manual Handi R218M12 001 Operation Manual Handi Inte...

Страница 15: ...calibration value Make sure that the calibration gas is connected to the Handi at 2 liters per minute for a minimum of 2 minutes prior to proceeding with calibration This 2 minute equilibration time...

Отзывы: