 6.0 CLEANING AND MAINTENANCE

When cleaning or disinfecting the Handi+, take appropriate care to prevent 
any solution from entering the analyzer.  
»  The Handi+ surface may be cleaned using a cloth moistened with 65% 

isopropyl alcohol/water solution or germicidal wipe.  

»  The Handi+ may be disinfected using standard topical disinfectants. 
»  The Handi+ is not intended for steam, ethylene oxide or radiation sterilization.
»  Store the Handi+ in a temperature similar to its ambient environment of daily use.

 7.0 SPECIFICATIONS

7.1 Analyzer Specifications

Measurement Range: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0.0 - 99.9% oxygen (gas)
Resolution/Display: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.1%  

The three digit LCD indicates values between 0.0 - 99.9% oxygen 

Over range indicated by one decimal point on display located after the first digit

Accuracy and Linearity:  . . . . . . . . . .1% of full scale at constant temperature, R.H. and
 

pressure when calibrated at full scale

Total Accuracy:. . . . . . . . . . . . . ±3% actual oxygen level over full operating temp range
Response Time: . . . . . . . . . . . . . . . . . . . .< 15 seconds for 90% step change. (at 25˚C)
Warm-up Time: 

None required

Operating Temperature:  . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15˚C - 40˚C (59˚F - 104˚F)
Storage Temperature: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .-15˚C - 50˚C (5˚F - 122˚F)
Atmospheric Pressure: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 800-1013 mBars 
Humidity: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-95%(non-condensing)
Environmental: . . . . . . . . . . . . . . . . . . General purpose housing equivalent to NEMA 1.

The 

Handi

+ is not waterproof.

Warranty: . . . . . . . . . . . . . . . . . . . . . . . . . Two (2) years in normal operating conditions
Power Requirements: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Powered by one internal, 

non-replaceable Lithium battery, CR2450.  

Power on push button automatically shuts off after 80 seconds time-out. 

Electronics rated general purpose; 

not for use in hazardous areas or for use with flammable gases.

Sensor Type:  . . . . . . . . . . . . . . . . . . . . . . . . .Maxtec MAX-250 series galvanic fuel cell
Weight:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Approx. 60 grams

Содержание Handi+ R218P12

Страница 1: ...H Handi O P E R A T I N G M A N U A L I N S T R U C T I O N S F O R U S E R 2 1 8 P 1 2 R218M12 Rev J CAL...

Страница 2: ......

Страница 3: ...s from the date of shipment from Maxtec under normal operating conditions and provided that the Handi is properly operated and maintained in accordance with Maxtec s operating instructions Should Hand...

Страница 4: ...Improper use of this device can cause inaccurate oxygen readings which can lead to improper treatment hypoxia or hyperoxia Follow the procedures outlined in this user manual CAUTION Indicates a potent...

Страница 5: ...ation A 3 2 4 Direct Flow Calibration Configuration B 4 3 0 OPERATING THE HANDI 4 4 0 FACTORS INFLUENCING ACCURATE READINGS 5 4 1 Elevation Changes 5 4 2 Temperature Effects 5 4 3 Pressure Effect 5 4...

Страница 6: ......

Страница 7: ...blank The analyzer turns off after 1 3 minutes if no buttons are pressed Calibration Button Press the calibration CAL button to adjust the calibration value to reflect the known oxygen concentration T...

Страница 8: ...his oxygen analyzer ideal for portable oxygen analysis by qualified heath care professionals 2 0 CALIBRATING THE Handi 2 1 Before You Begin A protective film covering the threaded sensor face must be...

Страница 9: ...encing Calibration and Performance on page 5 of this manual Note We recommend use of medical grade USP or 99 purity oxygen when calibrating the Handi 2 3 In Line Calibration Configuration A 1 Put the...

Страница 10: ...ess the CAL button on the Handi The calibration gas value on the analyzer display should read 20 9 or 100 depending on the gas stream used 3 0 OPERATING THE HANDI To Check the Oxygen Concentration of...

Страница 11: ...e Handi will hold calibration and read correctly within 3 when in thermal equilibrium within the operating temperature range The device must be thermally stable when calibrated and allowed to thermall...

Страница 12: ...an error code occurs E03 No valid calibration data available Make sure unit has reached thermal equilibrium Press and hold the Calibration Button for three seconds to manually force a new calibration...

Страница 13: ...point on display located after the first digit Accuracy and Linearity 1 of full scale at constant temperature R H and pressure when calibrated at full scale Total Accuracy 3 actual oxygen level over...

Страница 14: ...on Dioxide 10 2 Helium 70 2 8 0 HANDI SPARE PARTS AND ACCESSORIES 8 1 Standard Replacement Parts and Accessories Part Number Item R218M12 Operation Manual Handi R218M12 001 Operation Manual Handi Inte...

Страница 15: ...calibration value Make sure that the calibration gas is connected to the Handi at 2 liters per minute for a minimum of 2 minutes prior to proceeding with calibration This 2 minute equilibration time...

Отзывы: