 1.0 INTRODUCTION

1.1 Component Identification

LCD Display:

 

A 3-digit display provides a direct readout of oxygen 

concentration in the range of 0 - 100%. The display is blank when the Handi+ 
enters its Sleep (power off) mode. The Handi+ will automatically enter the 

Sleep mode after approximately 1.3 minutes from the last time the ON button 

was pressed.

ON/OFF Button:

 

Use this button to turn the Handi+ on and off. When the 

Handi+ is in the Sleep (power off) mode, the LCD display is blank. 

The analyzer turns off after 1.3 minutes if no buttons are pressed.

Calibration Button:

 

Press the calibration (CAL) button to adjust the calibration 

value to reflect the known oxygen concentration. To simplify operation, the 
Handi+ Analyzer automatically determines the calibration gas being used as 

room air (20.9%) or high grade (100%).

Flow Diverter:

 

The flow diverter is designed to fit industry standard 15 mm  

I.D. "T" adapters.

Over Range Indicator:

 

The appearance of a decimal point after the first digit 

means that the Handi+ is reading in excess of 99.9%.

Example:  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.0.0 = 100%   

  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.0.1 = 101%

  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.0.2 = 102%
  . . . . . . . (If the display reads > 0.0.3 the Handi+ should be re-calibrated.)

CALIBRATION BUTTON

LCD DISPLAY

"ON" BUTTON

            TOP VIEW

Содержание Handi+ R218P12

Страница 1: ...H Handi O P E R A T I N G M A N U A L I N S T R U C T I O N S F O R U S E R 2 1 8 P 1 2 R218M12 Rev J CAL...

Страница 2: ......

Страница 3: ...s from the date of shipment from Maxtec under normal operating conditions and provided that the Handi is properly operated and maintained in accordance with Maxtec s operating instructions Should Hand...

Страница 4: ...Improper use of this device can cause inaccurate oxygen readings which can lead to improper treatment hypoxia or hyperoxia Follow the procedures outlined in this user manual CAUTION Indicates a potent...

Страница 5: ...ation A 3 2 4 Direct Flow Calibration Configuration B 4 3 0 OPERATING THE HANDI 4 4 0 FACTORS INFLUENCING ACCURATE READINGS 5 4 1 Elevation Changes 5 4 2 Temperature Effects 5 4 3 Pressure Effect 5 4...

Страница 6: ......

Страница 7: ...blank The analyzer turns off after 1 3 minutes if no buttons are pressed Calibration Button Press the calibration CAL button to adjust the calibration value to reflect the known oxygen concentration T...

Страница 8: ...his oxygen analyzer ideal for portable oxygen analysis by qualified heath care professionals 2 0 CALIBRATING THE Handi 2 1 Before You Begin A protective film covering the threaded sensor face must be...

Страница 9: ...encing Calibration and Performance on page 5 of this manual Note We recommend use of medical grade USP or 99 purity oxygen when calibrating the Handi 2 3 In Line Calibration Configuration A 1 Put the...

Страница 10: ...ess the CAL button on the Handi The calibration gas value on the analyzer display should read 20 9 or 100 depending on the gas stream used 3 0 OPERATING THE HANDI To Check the Oxygen Concentration of...

Страница 11: ...e Handi will hold calibration and read correctly within 3 when in thermal equilibrium within the operating temperature range The device must be thermally stable when calibrated and allowed to thermall...

Страница 12: ...an error code occurs E03 No valid calibration data available Make sure unit has reached thermal equilibrium Press and hold the Calibration Button for three seconds to manually force a new calibration...

Страница 13: ...point on display located after the first digit Accuracy and Linearity 1 of full scale at constant temperature R H and pressure when calibrated at full scale Total Accuracy 3 actual oxygen level over...

Страница 14: ...on Dioxide 10 2 Helium 70 2 8 0 HANDI SPARE PARTS AND ACCESSORIES 8 1 Standard Replacement Parts and Accessories Part Number Item R218M12 Operation Manual Handi R218M12 001 Operation Manual Handi Inte...

Страница 15: ...calibration value Make sure that the calibration gas is connected to the Handi at 2 liters per minute for a minimum of 2 minutes prior to proceeding with calibration This 2 minute equilibration time...