![MAVIG CT Expres Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/mavig/ct-expres/ct-expres_service-manual_1742669027.webp)
Page 27 of 54
5100054
BND01O03.docx
Figure 21 Cable routing through Spring Arm
•
The spring arm has to be positioned at
an angle of 0° (horizontally).
3.
Take off the three cable clips
①
.
4.
Route the cables
③
towards the parallel
guide
②
at the bottom of the spring arm
and then route them through the spring
arm next to each other and without
twisting as illustrated in Figure 21.
•
Make sure you position the cable
with the largest cross section
centrally.
5.
Click the three cable clips
①
back onto
the parallel guide at their previous
position.
6.
Check the clearance of the cables
③
:
•
The cables
③
must be laid to the
front and rear joint of the spring
arm with a slight, strain-free
curvature.
•
Check that the external cables
③
are not exposed to tensile stress in
any possible position and that the
spring arm can be freely moved.
Figure 22 Cable routing through the Injector
Column
7.
Route the cables through the injector
column and out of the aperture
④
.
④