Troubleshooting
page 16/20
11. Troubleshooting
Relieve the pressure prior to checking or repairing the meter.
Make sure all valves, controls and pumps are operating correctly.
Symptom
Possible Cause
Remedy
Battery Icon is displayed
Batteries are low
Replace batteries
Display Blank
Meter asleep
Push reset button
Loose battery connection
Remove battery pack
and check battery connection
/ Push reset button
Batteries dead
Replace batteries / Push
reset button
Meter does not latch for
Meter not in AUTO mode
Press AUTO button and
batching
program batch size
Meter not reset after prior batch
Press RESET button
Low batteries
Check for battery icon
/ replace batteries / push
RESET button
Slow or no fluid flow
Filter is clogged
Clean or replace the filter in
the swivel nut
Pump pressure is low
Turn up the pump pressure
Foreign material is jamming meter
Contact your local repair
distributor for
Meter inaccurate
Scale factor not correct for fluid
Enter program mode,
check and reset
program factor
Batch overruns program value
Pulse delay value set too low
Enter program mode, reset
pulse delay to higher value