3-17
Masterflex REGLO Digital Pump Drive with Advanced Connectivity
Section 3: Operation
Masterflex
I. SCREEN LOCK: Locking the screen disables all touchscreen functions.
To lock the screen:
Press and
hold UNLOCKED until the icon changes to LOCKED and a red border appears around the
screen.
To unlock the screen:
Press and hold LOCKED until the icon changes to UNLOCKED .
J. SETTINGS: Tap to access the Settings screen (for further information see "
Continuous Mode Operation
General Preparation:
• Confirm the correct tubing has been selected and is properly loaded into the pump drive (for further
• Confirm all fluids and containers are ready.
• Switch the drive on using the power switch located at the rear of the drive.
NOTE:
The touchscreen
display will revert to the previously used operation mode.
• Confirm that the tubing has been calibrated by checking that Calibration Complete is displayed on the
Continuous Mode Run screen (for further information see "
• If required, prime the pump before operation (for further information see "
page 3-7).
To operate the pump in Continuous Mode:
1. Tap CONTINUOUS from the Mode Selection screen. The Continuous Mode Run screen will be
displayed.
2. If required, Press and hold CUMULATIVE VOLUME RESET to reset the display to zero.
3. Select the desired flow rate and flow units:
a. Tap FLOW RATE to access the Flow Rate screen.
b. Tap UNITS to access the Flow Units screen and select the desired flow unit from the available list.
c. Tap CONFIRM to save or CANCEL to discard changes and return to the Flow Rate screen.
d. Tap FLOW and enter the desired flow rate using the onscreen keypad.
e. Tap CONFIRM to save or CANCEL to discard changes and return to the Continuous
Mode screen.
4. Tap SIZE. The Tube Size screen will be displayed.
a. Select the desired tube size from the available list or
SCROLL
and tap CUSTOM to enter a
custom tube size using the onscreen keypad.
b. Tap CONFIRM to save or CANCEL to discard changes and return to the Continuous
Mode Run screen.
5. If required, tap CALIBRATE to calibrate the pump for the selected tube and flow rate.
6. Tap DIRECTION to select either clockwise
or counterclockwise
flow direction.
7. Tap START . The drive will commence operation at the flow rate and direction shown.
8. Tap STOP when the drive operation is no longer required.
NOTE:
In Continuous Mode the drive
will continue to operate at the displayed flow rate and direction until stopped.