3-8
Masterflex REGLO Digital Pump Drive with Advanced Connectivity
Section 2: Basic Setup & Settings
Masterflex
TUBE CALIBRATION
Tube calibration should be performed whenever fluids, flow rate or tubing is changed to ensure accurate
dispensing of fluids. Tube calibration can be accessed from the Continuous Mode Run screen or from either
of the Volume or Time Mode Edit screens.
Tube Calibration Screen
General Preparation:
• Confirm the correct tubing has been selected and is properly loaded into the pump drive (for further
• Confirm all fluids and containers are ready.
To calibrate tubing:
1. Insert the tube inlet into the supply fluid.
2. Insert the tube outlet into a suitable container.
NOTE:
The container should be graduated or placed on
a scale for increased accuracy. If using a scale, an acceptable weight to volume conversion for water is: 1
gram = 1 mL.
3. Switch the drive on using the power switch located at the rear of the drive.
4. Navigate to the Continuous, Time or Volume Mode screens.
5. Tap DIRECTION to select either clockwise
or counterclockwise
flow direction.
6. If required, Tap START and allow the pump to run for approximately 20-30 minutes. This step is
recommended to properly condition new tubing and to increase calibration accuracy.
NOTE:
If required,
the pump can be run dry while conditioning the tube.
7. Press and hold PRIME to prime the pump. Priming will stop when PRIME is released.
Selected Tube Size
Approximate Duration
Set Flow Rate
Calibration Volume
Estimated Performance
Start Calibration