SYSTEM BOOT—
EXTERNAL POWER
Open the exterior housing, connect the AC/DC Universal
Power Supply/Converter into the EXT PWR IN receptacle on
the front. Then connect the AC plug into an AC receptacle.
The system will come on automatically. The first screen dis-
played on the LCD panel will be:
This will be followed by another screen:
Remove the plug from the EXT PWR IN receptacle on the
control console. LCD will change to read:
The system will power up the drive system to place the
internal battery under load and verify its condition. After the
drive system has stopped, you may reconnect the external
power to the EXT PWR IN receptacle. The LCD panel will
display one of the following messages, depending upon
battery condition.
If the battery is fully charged or you pressed cancel when
notified the battery is low, the system will complete boot-up
and display the following on the LCD screen:
If the message displayed is:
The system should not be operated until the battery has been
charged. Proceed to the BATTERY CHARGING section of the
manual on page 10.
SAMPLER ON VERSION X.X
REMOVE EXT PWR TO CHECK
BATTERY—CANCEL TO SKIP
CHECKING BATTERY
PRESS CANCEL TO QUIT
BATTERY IS FULLY CHARGED
BATTERY IS LOW
CHARGING IS RECOMMENDED
BATTERY IS VERY LOW
MUST CHARGE BATTERY!
SAMPLER SYSTEM READY
HH:MM:SS
MM/DD/YYYY
9